Owner's manual

YSI 5200A
121
YSI 5200A
120
1. Enter Name Devices/Sensors menu
Menus → System → Name Devices/Sensors
2.
Use
and
to scroll and highlight submenus.
3.
Press
to select.
4. Use keypad
to congure name.
5. Press
to save conguration and return to Run Screen.
Notes:
- Instrument, Aux, and Sensor Names are not cleared at Reset to Factory
menus.
- Relay names are cleared at “reset relays, “reset all systems, and “reset to
factory/reset all” Service menus.
- e 5200A default system name is displayed at Event, Sensor, and Calibra-
tion logs, regardless if name is user congured.
- Press “Clear” at keypad and then press
to clear user congured
names.
- Aux system names can also be congured at: Menus → System → Con-
gure Aux →Aux 1-2→Probe Name
- Because of font limitations, 1x2 and 2x3 display formats do not show all
20 characters of a 20 character name. Typically only 13 characters are
displayed.
- e instrument name is not shown at Run Screen. e instrument name
is shown at relay enable menus.
- Certain menus (i.e. Congure Aux) display up to 18 characters of the de-
vice name. “...” is shown at the menu if the name exceeds 18 characters.
Password
Menus → System → Password
Enable and set the password in the Password menu. Once the password function
is enabled and saved, the 5200A system menus are not accessible without rst
Conguring the 5200A
entering the correct password at the Run Screen. e password can contain up to
12 characters.
1. Enter Password menu
Menus → System → Password
2.
Use
and
to scroll and highlight submenus.
3.
Press
to select.
4. Enable and congure submenus.
5. Press
to save conguration and return to Run Screen.
Notes:
-
is not password protected.
-
icon displays at Run Screen when password function is enabled.
Power Fail Menu
Menus → System → Power Fail
In the event of AC power failure, an AC version instrument’s power switches to
DC if properly wired. e switch from AC to DC power is automatic and does not
require 5200A conguration. Enable Power Fail AC Monitor to generate event when
AC power fails and is restored.