Owner's manual
YSI 5200A
115
YSI 5200A
114
6. AquaManager Program Folder includes the latest version of instru-
ment rmware. e lastest rmware is stored as an .mdb le in the PC
Program Files>AquaManager>Firmware. At AquaManager, Firmware
Updater>File>Load Firmware Database. Locate and load this le into the
Firmware Updater.
7. Le click
button.
window opens.
8. Select “Yes” and the following screens will appear.
On instrument On PC
YSI 5200A/5400/5500 Bootstrap Loader screen Flash Programmer uploading rmware into
5200A
Conguring the 5200A
Note:
- If the YSI 5200A/5400 Bootstrap Loader screen does not appear at the
instrument aer above steps manually place 5200A in downloader at
Menus → System → Downloader. Close and reopen AquaManager. Reopen
Flash Programmer and repeat steps starting at 6.
- Power cycle to exit downloader and return to run screen. Do not power
cycle during an active upload.
- If aer manually placing the 5200A into downloader and the Flash Pro-
grammer utility still does not upload rmware see Force Downloader
- page 116.
9. e 5200A will reboot (power cycle) when rmware upload is complete.
10. Select “ok” when this window appears
11. Close Flash Programmer.
12. Verify version information to conrm downloaded code:
13. Menus → System → Version
14. If applicable, disable Force Serial Port conguration - page 100.
Notes:
- During upload, 5200A systems are placed into Service Hold which de-
energizes any active systems including energized relays. See SVR Hold
Menu - page 169 for additional information. No “o” event is created
for system relays that de-energize when 5200A is placed in downloader.
Systems are reset when power is cycled aer downloader.
- All logs and instrument conguration are retained when 5200A rmware
is updated.
- Display board code loads rst, followed by control board code. e display
board screen ickers when display board code is loading. e display stops
ickering when the control board code is loading. e Firmware Updater
progress bar will complete two full scrolls during rmware updating.
- Total download time is approximately nine minutes; eight minutes to
Conguring the 5200A