Owner's manual

YSI 5200A
113
YSI 5200A
112
1. Enter Display Settings menu
Menus → System → Display Settings →Display Format
2.
Use
and
to scroll and highlight submenus.
3.
Press
to select.
4. Press
to save conguration and return to Run Screen.
Notes:
- At 1x2 & 2x2 display settings, analog aux values over 7 digits will overlap
on to the adjoining sensor window. Adjust aux display format to lower
resolution.
- Because of font limitations, 1X2 and 2X3 display formats do not show
all 20 characters of a 20 character name. Typically only 13 characters
are displayed.
Updating 5200A Firmware
Like all microprocessor based instruments, it may be necessary to upload revisions
to the 5200A rmware. Download new rmware using the Firmware Updater utility
in AquaManager. A serial RS232 connection is necessary to run the Downloader
program.
e instrument rmware may become corrupted if any of the following occurs
during a 5200A rmware update:
- Firmware Updater utility is closed;
- AquaManager is closed;
- RS232 cable is removed from 5200A or PC com port;
- power is disconnected from 5200A.
If the 5200As rmware becomes corrupted and is not updated properly, the instru-
ment will need to be returned to YSI.
Conguring the 5200A
Firmware Updater Directions
1. Wire RS232 connection. See RS232 Communication - page 59. (Dis-
regard if already using RS232 connection.)
Note:
Connect RS232 directly into slaves in order to upload rmware to slaves.
Disable slaves at the master to avoid slave failure alarms when uploading
- Menus → System → Communications → Network →Master.
2. Enable the Force Serial Port on 5200As wired and congured with an
ethernet device - page 100.
3. Verify that the baud rate in 5200A Communications menu is congured
to 115200.
4. From the AquaManager Explorer window, access the Firmware Updater
utility from the pull down menu: Tools Instrument toolsUpdate
Firmware.
5. Congure PC com port number (1-8) in Flash Programmer.
Notes:
- Use only a Tripp-Lite Keyspan USB-Serial Adapter, model USA-19HS
YSI p/n 773.
- When using USB-Serial Adapter, verify that the port number assigned
during driver installation is the same port congured when using Aqua-
Manager Flash Programmer.
Conguring the 5200A