- Use AquaManager to store and analyze 5200A logs. To avoid gaps in data
use polling intervals based on the user dened data logging conguration.
Set the data log interval so that analyzing the data is meaningful.
- AquaManager users can view Configuration events in the
Conguration log. Conguration
events are created when certain 5200A
conguration changes are made. e conguration log is not accessible
at the 5200A. See AquaManager Online Help for additional information.
Data Logging
Menus → System →Data Logging→Data Logging
e 5200A sensor log stores records for sensors and aux systems. Sensor logs are
time and date stamped based on the user congured logging interval. e logging
interval is user congured at 1-255 seconds, minutes or hours.
1. Enter Data Logging menu
Menus → System →Data Logging→Data Logging
2.
Use
and
to scroll and highlight submenus.
3.
Press
to select.
4. Enable and congure submenus.
5. Press
to save conguration and return to Run Screen.
Press
to view the 25 most recent sensor logs.
Notes:
- Temperature values are stored in °C regardless of display format.
- Aux digital “closed” is logged as Aux 0.00. Aux digital “open” is logged
as Aux 1.00.
- e 5200A sensor log records 0.00 resolution for aux analog inputs regard-
less of the selected display format.
- DO values are stored in mg/L & %Sat regardless of selected display
format.
Conguring the 5200A
- Sensor logs record the default system name and not the user congured
system name.
- Changes to the data logging interval take eect aer the expiration of the
interval being changed. Example - if the interval is changed from one
hour to one minute, one minute logging will begin aer the expiration of
the current one hour interval. Power cycle the 5200A to make a change
to the data logging interval occur immediately.
- Menu congurations are saved to the system by pressing
to
return to the Run Screen and not when the change is made in the menu.
Exception - Date and time and clear data logs are updated immediately
to the system, not when
is pressed.
- e sensor log contains values for all sensor and aux systems regardless
if the system is disabled.
- 5200A sensor log values for disabled systems and systems without sen-
sors are not valid.
- AquaManager data log displays 5200A conductivity values in µS to 0.00
resolution.
In the above sensor log screen shot ORP is disabled.
Event Logging
Menus → System → Data Logging → Event Logging
Control, Alarm, and Timer relays assigned to 5200A systems create events when
they turn on and o. In addition, AC power failures, ethernet device error messages,
and slave network failures are create events. ese events are stored in the Event
Log. Event logs identify the 5200A system, are time and date stamped and record
relay the action (“on” or “o”).
1. Enter Event Logging menu
Menus → System → Data Logging → Event Logging
2.
Use
and
to scroll and highlight submenus.
Conguring the 5200A