- Use AquaManager to store and analyze 5200A logs. To avoid gaps in data
use polling intervals based on the user dened data logging conguration.
Set the data log interval so that analyzing the data is meaningful.
- AquaManager users can view Configuration events in the
Conguration log. Conguration
events are created when certain 5200A
conguration changes are made. e conguration log is not accessible
at the 5200A. See AquaManager Online Help for additional information.
Data Logging
Menus → System →Data Logging→Data Logging
e 5200A sensor log stores records for sensors and aux systems. Sensor logs are
time and date stamped based on the user congured logging interval. e logging
interval is user congured at 1-255 seconds, minutes or hours.
1. Enter Data Logging menu
Menus → System →Data Logging→Data Logging
2.
Use
▲
and
▼
to scroll and highlight submenus.
3.
Press
to select.
4. Enable and congure submenus.
5. Press
to save conguration and return to Run Screen.
Press
to view the 25 most recent sensor logs.
Notes:
- Temperature values are stored in °C regardless of display format.
- Aux digital “closed” is logged as Aux 0.00. Aux digital “open” is logged
as Aux 1.00.
- e 5200A sensor log records 0.00 resolution for aux analog inputs regard-
less of the selected display format.
- DO values are stored in mg/L & %Sat regardless of selected display
format.
Conguring the 5200A
- Sensor logs record the default system name and not the user congured
system name.
- Changes to the data logging interval take eect aer the expiration of the
interval being changed. Example - if the interval is changed from one
hour to one minute, one minute logging will begin aer the expiration of
the current one hour interval. Power cycle the 5200A to make a change
to the data logging interval occur immediately.
- Menu congurations are saved to the system by pressing
to
return to the Run Screen and not when the change is made in the menu.
Exception - Date and time and clear data logs are updated immediately
to the system, not when
is pressed.
- e sensor log contains values for all sensor and aux systems regardless
if the system is disabled.
- 5200A sensor log values for disabled systems and systems without sen-
sors are not valid.
- AquaManager data log displays 5200A conductivity values in µS to 0.00
resolution.
In the above sensor log screen shot ORP is disabled.
Event Logging
Menus → System → Data Logging → Event Logging
Control, Alarm, and Timer relays assigned to 5200A systems create events when
they turn on and o. In addition, AC power failures, ethernet device error messages,
and slave network failures are create events. ese events are stored in the Event
Log. Event logs identify the 5200A system, are time and date stamped and record
relay the action (“on” or “o”).
1. Enter Event Logging menu
Menus → System → Data Logging → Event Logging
2.
Use
▲
and
▼
to scroll and highlight submenus.
Conguring the 5200A