Owner's manual

YSI 5200A
105
YSI 5200A
104
Input Scale Min and Max values
Menus → System → Congure Aux →Aux 1-2 →Input Scale Min
→Input Scale Max
Input scaling allows a way to view data withn a specic input range of the input type
(0-1v, 0-5v, 420/ma) in greater resolution. In the below example 0-1v is the input
type and display format = 0. e congured Min value is 0, and the congured Max
value is 1000. e input scale Min is .25 and the input scale Max is .75. (e user
wants to only view data within this specic range). In the below example an input
value of .25v = a display and datalog value of 0. An input value of .50v = a display
value and datalog value of 1000.
Display Format
Menus → System → Congure Aux →Aux 1-2→Display Format
Congure the Display format (0, 0.0, or 0.00) based on the resolution to be displayed
on the Run Screen.
Example - when using a total dissolved gas pressure sensor congure, the display
format as 0.0 - gure 4.6, page 105.
In the below screen shot, the Run Screen is displaying aux name “TGP,” sux
mmHg,” with display format “0.0”. e display value represents an input of 1.875
volts with 0-5VDC analog aux conguration. e min value = “-400”. e max
value = “+400”. e control/alarm low value is “-50”. Alarms are enabled for this
aux input.
Conguring the 5200A
gure 4.6
Notes:
- Values are displayed based on the display format resolution. Example - if
the display format is 0.0 and the min value congured is 1.88, then 1.9
will be the value displayed in the min value menu.
- 5200A aux sensor logs record 0.00 resolution regardless of the selected
resolution in the aux analog display format conguration.
1. Wire aux analog devices - page 49.
2. Enter Aux menu
Menus → System → Congure Aux →Aux 1-2
3.
Use
and
to scroll and highlight submenus.
4.
Press
to select.
5. Press
to save conguration.
6. Go to Sensor Setup Menus  Sensor Setup Aux 1-2 - page 142 to enable
and congure set points, control, alarm, relays, and hold o time.
7. Press
to save conguration.
Data Logging
Menus → System →Data Logging
e 5200A saves information in the calibration, sensor, and event logs. Press
at the Run Screen to view the 25 most recent records for each log. e 5200A sen-
sor log stores 2000 records. e calibration log can store 500 records. e event
log can store 1250 records. Once a 5200A log is full, it will begin overwriting the
oldest records with new records (data wrapping).
Conguring the 5200A