Owner's manual
YSI 5200A
105
YSI 5200A
104
Input Scale Min and Max values
Menus → System → Congure Aux →Aux 1-2 →Input Scale Min
→Input Scale Max
Input scaling allows a way to view data withn a specic input range of the input type
(0-1v, 0-5v, 420/ma) in greater resolution. In the below example 0-1v is the input
type and display format = 0. e congured Min value is 0, and the congured Max
value is 1000. e input scale Min is .25 and the input scale Max is .75. (e user
wants to only view data within this specic range). In the below example an input
value of .25v = a display and datalog value of 0. An input value of .50v = a display
value and datalog value of 1000.
Display Format
Menus → System → Congure Aux →Aux 1-2→Display Format
Congure the Display format (0, 0.0, or 0.00) based on the resolution to be displayed
on the Run Screen.
Example - when using a total dissolved gas pressure sensor congure, the display
format as 0.0 - gure 4.6, page 105.
In the below screen shot, the Run Screen is displaying aux name “TGP,” sux
“mmHg,” with display format “0.0”. e display value represents an input of 1.875
volts with 0-5VDC analog aux conguration. e min value = “-400”. e max
value = “+400”. e control/alarm low value is “-50”. Alarms are enabled for this
aux input.
Conguring the 5200A
gure 4.6
Notes:
- Values are displayed based on the display format resolution. Example - if
the display format is 0.0 and the min value congured is 1.88, then 1.9
will be the value displayed in the min value menu.
- 5200A aux sensor logs record 0.00 resolution regardless of the selected
resolution in the aux analog display format conguration.
1. Wire aux analog devices - page 49.
2. Enter Aux menu
Menus → System → Congure Aux →Aux 1-2
3.
Use
▲
and
▼
to scroll and highlight submenus.
4.
Press
to select.
5. Press
to save conguration.
6. Go to Sensor Setup Menus Sensor Setup Aux 1-2 - page 142 to enable
and congure set points, control, alarm, relays, and hold o time.
7. Press
to save conguration.
Data Logging
Menus → System →Data Logging
e 5200A saves information in the calibration, sensor, and event logs. Press
at the Run Screen to view the 25 most recent records for each log. e 5200A sen-
sor log stores 2000 records. e calibration log can store 500 records. e event
log can store 1250 records. Once a 5200A log is full, it will begin overwriting the
oldest records with new records (data wrapping).
Conguring the 5200A