Owner's manual
YSI 5200A
103
YSI 5200A
102
Notes:
- Sux, Min, Max, Input Scale Min, Input Scale Max and Display Format
are not available for digital inputs.
- In the 5200A sensor log, aux digital “closed” is logged as Aux 0.00. Aux
digital “open” is logged as Aux 1.00.
1. Wire aux digital devices - page 49.
2. Enter Aux menu
Menus → System → Congure Aux →Aux 1-2
3.
Use
▲
and
▼
to scroll and highlight submenus.
4.
Press
to select.
5. Congure digital input type and name.
6. Press
to save conguration and return to Run Screen.
7. Go to Sensor Setup Menus Sensor Setup Aux 1-2 - page 142 to enable
and congure control/alarm on condition, control/alarm relay, and hold
o time.
8. Press
to save conguration.
Analog 0-1VDC and 0-5VDC and 4-20 mA
Menus → System → Congure Aux →Aux 1-2→Input Type →0-1V
→0-5V
→4/20mA
Select the analog input type based on the signal being sent from the device.
Notes:
- It is normal to hear an on-board relay “click” when switching to or from
the 4-20mA aux input type.
- Analog systems will not monitor, control, or alarm properly if a device
that operates outside the congured analog range is installed and con-
gured.
Conguring the 5200A
Sufx
Menus → System → Congure Aux →Aux 1-2→Sux
Enter the Sux (up to 4 characters) to describe the input type or units. e sux
is displayed at the Run Screen. e sux is typically used to describe the input
signal type.
Example - when using a total dissolved gas pressure sensor, “mmHg” could be used
as a sux for measuring TGP (total dissolved gas pressure) - gure 4.6, page 105.
Min and Max values
Menus → System → Congure Aux →Aux 1-2 →Min
→Max
Min and Max values are user dened. e range is -10000 to 10000. Min and Max
values are used for scaling the input value received from an analog aux input. e
min and max values may be congured to correspond to the minimum input level
(i.e. 0VDC or 4mA) and maximum input level (i.e. 1VDC, 5VDC, or 20mA). e
Min value must be less than Max value.
Example - when using a total dissolved gas pressure sensor (0-5VDC input type)
enter -400 for min value. e 5400 will display “-400” when the input value is 0VDC.
Set the max value to correspond to the maximum input value as +400. e 5200A
will display “400” when input value is 5VDC.
Notes:
- e 5200A will monitor and data log based on the “min” and “max” volt-
age or current signal.
- To enter negative values, type the whole number then select “-” on the
numeric keypad.
- Whenever min and/or max values are changed, go to Menus → System
→ Sensor Setup → Aux 1-2 → Control Low and Control High menu to
verify control/alarm values are correct for the application.
Conguring the 5200A