Owner's manual

YSI 5200A
101
YSI 5200A
100
Force Serial Port
Menus → System → Communications → Force Serial Port
Enable the Force Serial Port when the 5200A is communicating via RS232 (direct)
to a PC and when the 5200A has a serial to ethernet device.
1. Wire RS232 connection - page 59
2. Enter Network menu
Menus → System → Communications → Force Serial Port
3. Use the
sokeys to congure o or on.
4. Press
to save conguration and return to Run Screen.
Notes:
- RS232 connection is required to upload rmware using the Firmware
Updater.
- e RS232 icon
blinks at the Run Screen when in Force Serial Port
mode.
Non-Networked (Stand Alone) 5200As
Non-networked (stand alone) 5200As must have the termination relay enabled. In
addition, non-networked (stand alone) 5200As must be congured as a master in
the System Communication menu with no slave nodes enabled.
Congure AUX
Menus → System → Congure Aux 1-2
e 5200A has two auxiliary (Aux) inputs. Aux inputs can be congured as digital or
analog (0-1VDC, 0-5VDC, or 4-20 mA) inputs. Congure sux, minimum (Min),
maximum (Max), Input Scale Min, Input Scale Max and display format menus for
analog inputs. Go to Sensor Setup Menus  Sensor Setup Aux 1-2 - page 142 to
enable Aux system and congure control/alarm settings. Both digital and analog
directions follow. Submenu functionality is provided aer conguration directions.
Conguring the 5200A
Probe Name
Menus → System → Congure Aux →Aux 1-2→Probe Name
Use the alpha/numeric keypad to name aux inputs. Congured name(s) appear
throughout menus and at Run Screen - gure 4.6, page 105.
Note:
- Aux names can also be congured at Menus  System  Name Devices/
Sensors.
Input Type
Menus → System → Congure Aux →Aux 1-2→Input Type
Select the input type for the device(s) wired to aux system(s). Aux Sensor Setup (page
142) menu selections are based on the aux input type congured at the System menu.
Digital
Menus → System → Congure Aux →Aux 1-2→Input Type→Digital
Use the digital input when wiring an external device to control or alarm when the
contacts of the external switch are open or close. A common application would be
to detect high or low water levels using a oat switch.
Conguring the 5200A