Owner's manual
YSI 5200A
99
YSI 5200A
98
- Conguring a 5200A as a Master overrides any slave conguration that
had been previously saved.
- Slave alarms (slave alarm at master and 5200A alarm emails) will not oc-
cur if slave node serial number is not recognized at master node at steps
10. A or 10. B above.
Termination Relay
Menus → System → Communications → Network →Termination Relay
e termination relay must be enabled on the rst and last 5200A on a network. All
other devices should have the termination relay disabled.
Send Test Email Msg
Menus → System → Communications → Send Test Email Msg
Send a test email to verify proper conguration of ethernet menus. Test emails are
only transmitted from congured master 5200As.
Note:
- Test emails are sent to all congured email address.
- Exit to Run Screen is required before the test email is sent.
Example of a 5200A test email sent from an ethernet device:
Subject Line: 02/04/10 03:50:05 PM YSI5200A Email Alarm Test Message
From: myname@xxx-xxxx.com
Sent: Wed, Feb 04, 2010 3:49 pm
To: myname@xxx-xxxx.com
CC: sweet49@xxxx.com
Conguring the 5200A
Email Retry Delay
Menus → System → Communications → Email Retry Delay
Congure the Email Retry Delay for the time interval that emails should be retrans-
mitted aer the initial alarm email is sent. e initial alarm email is sent immediately
aer the general alarm hold o time or aer the aux hold o time. Emails will be
retransmitted until one of the bottom front panel sokeys is pressed, regardless if
the alarm condition still exists.
Note:
- On a network, emails are retransmitted according to the slave retry delay
conguration of the slave in alarm and not the master’s email retry delay
conguration. Please note, a slave Email Retry Delay menu will display
(Not available) because of its slave conguration; however, this feature
is available and has a default value of 15 minutes that can be congured
by the user.
- e Retry Delay should be congured at intervals ≥ 15 minutes.
- AquaManager alarm emails are retransmitted based on the 5200A slave(s)
and master email retry delay conguration.
Baud Rate RS232
Menus → System → Communications → Baud Rate
Congure the Baud rate (for RS232 connections only) in the 5200A to match the
baud rate congured in AquaManager. Use the default rate of 115200 for fastest
communication speed.
1. Enter Network menu
Menus → System → Communications → Network →Baud Rate
2.
Press
▲
or
▼
to scroll and highlight a selection.
3.
Press
to conrm selection.
4. Press
to save conguration and return to Run Screen.
Conguring the 5200A