Owner's manual
YSI 5200A
97
YSI 5200A
96
1. Wire the 5200A network - page 53.
2. Enter Network menu
Menus → System → Communications → Network
3.
Press
▲
and
▼
to scroll and highlight submenus.
4. For master node: Press
at master submenu.
5. For all slave nodes: Press
at slave submenu. Congure a slave address.
Notes:
- e slave address appears as the slave subnode at node properties
- Congure slave addresses with numbers of 1 through 31. Slave ad-
dresses of 32 to 64 will be used in future releases and should not be
congured with the current release.
6. Disable the Termination Relay on all devices except for the rst and last
device on the network. Menus → System → Communications → Network
→ Termination Relay.
7. Verify that the Termination Relay is enabled on the rst and last device
on the network.
8. For master node: Press
at master submenu. At assign slaves menu,
select the slave address congured in step 5 above. Press enter. Enable
one slave at a time using
sokeys . See Testing Network
- page 58 for the required testing network wiring.
Conguring the 5200A
9. Press to save conguration and return to Run Screen.
10. Each slave must be tested one by one. Insert network termination strip
and apply power to the master and the last physical device on the network.
If the master is not the rst or last physical device on the network, apply
power and insert network terminal strip on the master and on the rst
and last physical devices on the network. Refer to the network wiring
instructions for detailed testing information beginning on page 53.
Testing for proper wiring and conguration can be done one of two ways:
A. Aer performing step 9, wait at least 30 seconds and then reenter the Com-
munication Master submenu. If the serial number of the slave is displayed
next to the slave address number, the communication network is wired and
congured correctly.
OR
B. Aer performing step 9, wait at least 30 seconds and then poll the slave
node using AquaManager. If the serial number of the slave is displayed next
to slave address number, the network is wired and congured correctly.
11. Apply power. Insert network termination strip to 5200A/5400 units as
described in the network wiring instructions beginning on page 53.
Repeat network conguration directions starting at step 8 until network
wiring for master and all slaves has been tested.
Notes:
- An Event and General alarm (if enabled) will occur if, aer ten queries,
an enabled slave does not respond to the Master.
Conguring the 5200A