Owner's manual
YSI 5200A
95
YSI 5200A
94
Email Addresses
Menus → System → Communications → Ethernet → Email Address
→ Email Address
→ Email Address
→ CC/Email Address
Up to four email addresses can be congured for 5200A emails. At a minimum,
the rst email address and the CC/email address must be congured. If the CC/
email address is not congured, 5200A alarm emails will not contain any alarm
information. e email ‘from’ address will be the email address entered in the rst
Email Address menu.
Notes:
- See Send Test Email Msg - page 98 to verify proper conguration of
ethernet menus.
- On a network, both master and slave alarm emails are sent to the cong-
ured email addresses in the master.
- Test the alarm email delivery by creating an alarm condition at the 5200A
and verify the email(s) is received.
- e CC/Email address must be congured when using 5200A email alarm-
ing. Specic alarm system information will not be sent if the CC/Email
address menu is not congured with a valid email address.
- Use AquaManager to easily congure email addresses using the PC key-
board. See AquaManager on line help.
- If Autopoller is being used for AquaManager email alarming on a TCP/
IP connection, it is strongly recommended that the 5200A email alarming
using the serial to ethernet device be disabled. When the ethernet device
sends alarm emails, all Autopoller SMS communication is suspended.
is could potentially delay any AquaManager alarm emails from being
sent.
- To disable the serial to ethernet device from sending alarms, clear all four
email addresses from the Communication → Ethernet menus. Clear the
email addresses by using the “Spc” (space) key on the numeric keypad.
Save change at numeric keypad and press
to save conguration
and return to Run Screen.
Ethernet Module Error Message
At power up, the serial to ethernet device will generate an error message if it cannot
properly handshake with the 5200A hardware. If this error occurs, communication
(AquaManager and 5200A alarm emails) will not work via TCP/IP. e error could
be a result of incorrect serial to ethernet device installation or a faulty ethernet device.
Conguring the 5200A
e error message will appear in the event log.
SMS Messaging with Ethernet Module
SMS (Short Message Service) or text messaging is a service for sending messages to
your cell phone or mobile device. To use SMS messaging, enter the email address
in the Menus → System → Communications→ Ethernet →Email address. See Ap-
pendix 6 SMS - page 238 for a list of cellular and mobile companies that support
sending text messages and format information.
Node Network
Menus → System → Communications → Network
A node network refers to either one stand alone master node and/or one master
node and up to 31 congured and wired slave nodes. Node networks with at least
one slave are physically connected using RS485 protocol see page 53 for additional
information. Only the master node is congured for serial or ethernet module
communication.
On a network with at least one slave the master continuously queries enabled slaves
on the network. e master is the access point for the entire network. is includes
AquaManager communication and sending of 5200A/5X00 alarm emails. Slave
nodes must be congured with a slave address.
Notes:
- Multiple node networks communicating via direct connect (RS232) re-
quire separate PC COM ports for simultaneous AquaManager mapping
and autopolling.
Conguring the 5200A