Owner's manual

YSI 5200A
39
YSI 5200A
38
diameter. Some pre-drilled rubber grommets are supplied with the 5200A.
Figure 3.13
Accessing the I/O Board
e I/O board is accessible when the 5200A front panel is removed. e letters
on the metal plate, which is attached to the lower board assembly, identify ground
wire termination locations and pluggable terminal strip locations on the I/O board.
Letters on the metal plate correspond to the wiring diagram located on the inside of
the front panel - gure 3.14. A ribbon cable connects the display board (mounted
in the front panel) to the I/O board. When the 5200A front panel is removed, a
lanyard protects against stress on the ribbon cable. See Open Front Panel to Access
I/O Board - page 39.
Installation and Wiring
Figure 3.14
Open Front Panel to Access I/O Board
WARNING: Disconnect external power to the unit before doing any wir-
ing.
AVERTISSEMENT: déconnectez l’alimentation externe de l’unité avant
douvrir le panneau avant.
1. Loosen the 4 mounting screws connecting the front panel to the back
enclosure. Captive screws remain in front panel aer loosening.
2. Carefully li the 5200A front panel from the back enclosure. Front panel
should be placed to the le of enclosure. Allow the lanyard to hold the
weight of the front panel. e rubber gasket seated in the front panel
should remain in the front panel lid.
Installation and Wiring