Owner's manual

YSI 5200A
35
YSI 5200A
34
to these measurements. erefore, it is essential to have accurate measurements
at the sensors’ location. Locate the sensors of the probe assembly according to the
following specications.
Place the sensor end of the probe assembly:
• in a location that represents the entire system
• where the water ow is at least 3 inches per second (7.6 cm/sec) if using
the 2.0 mil PE membrane or at least 6 inches per second (15.24 cm/sec)
for 2.0 mil Teon membrane
• in a location that is free from debris (e.g. algae); ideally at a 45˚ angle from
vertical position and directed with the ow - gure 3.10
Figure 3.10
• in a well mixed, free-owing area; midstream and mid-depth typically gives
representative readings; the probe assembly and cable are fully submersible
up to the cable connector
• submersed at all times so liquid level uctuations do not expose the sen-
sors to the atmosphere
• so probe cabling is not routed near rotating machinery and/or equipment
involving electrical switching or regulation
• so that the probe assembly and cabling are located away from sources of
electrical interference such as UV sterilizers, orescent lighting, ballasts,
pumps, any high voltage peripheral devices, etc.
CAUTION: run high and low voltage cabling through separate bulkhead tting
and conduit
• where regular maintenance, including calibration can easily be done.
Calibrations are performed using the 5200A front panel.
Installation and Wiring
Note:
- All YSI sensors, except the temperature sensor on the 5561 cable, are eld
replaceable. Individual sensors can be ordered and replaced by the user
without returning the probe/cable assembly to the factory.
YSI 5561 Probe Assembly
e YSI 5561 probe assembly includes a DO and temperature sensor. e DO sensor
is shipped with a dry membrane. A new membrane with fresh electrolyte must be
installed on the DO sensor prior to use. Complete instructions are provided with
the sensor and membrane kits.
Notes:
- Replace the dry, shipping cap membrane with a 2 mil PE (blue) membrane
or 2.0 mil Teon (gray) membrane following the instructions provided
with the membrane kit.
- Do not discard the calibration/storage cup. It is used for calibrating and
storing the sensor when not in use.
- Allow approximately 10 to 15 minutes aer the 5200A is powered on, or
aer servicing the DO sensor, for the polarographic sensor to stabilize.
Figure 3.11 shows the external parts of the YSI 5561 probe assembly.
Figure 3.11
Installation and Wiring