YSI Model 3200 Conductance, Resistance, Salinity, Total Dissolved Solids and Temperature Instrument Operations Manual
CONTENTS 1. Introduction .............................................................................................................................................1 1.1 Capabilities .......................................................................................................................................1 1.2 User Features ....................................................................................................................................2 1.3 Controls..........................
7. Troubleshooting.....................................................................................................................................39 Error Messages ...............................................................................................................................39 8. Principles of Operation .........................................................................................................................41 8.1 Ratiometric Conductivity..................................
SECTION 1 INTRODUCTION The Model 3200 is a microprocessor based; menu-driven instrument designed to perform laboratory measurement of conductivity, resistivity, salinity, total dissolved solids and temperature. The instrument’s menu system makes it simple to use. The new YSI 3200 series conductivity cells contain internal temperature sensors for temperature measurement and automatic temperature compensation.
Introduction 1.2 Section 1 USER FEATURES • Audio feedback • Soft key/menu system • Real time clock • Auto linear temperature compensation calibration Two temperature points are tested in a known solution to automatically determine the solution’s compensation slope. Alternatively, the user may input a temperature compensation factor directly (0 - 10%/°C).
Introduction 1.3 Section 1 CONTROLS FRONT PANEL The front panel of the instrument contains the display and keypad as shown below. 3200 CONDUCTIVITY INSTRUMENT OPERATION 1.372uS 12:40PM CELL Display 24.
Introduction Section 1 The 3200 has two main operating modes, Data and Operation. The current mode is displayed in the dark bar at the top right corner of the display. The following diagram shows the typical display while the instrument is in the Operation mode. Conductance: S mS uS Conductivity: S/cm mS/cm uS/cm Resistance: Ω KΩ MΩ Resistivity: Ωcm KΩcm MΩcm Salinity: ppt TDS: mg/L Temperature: °C Mode Operation 1.953 S K=10.100 25.0°C u /cm 11:33AM 24.38°C 1.
Introduction Section 1 REAR PANEL The rear panel contains the connections for the power supply, cell, external temperature probe and RS232 serial port. It also has a mounting post for the 3232 adapter as shown below. Adapter mounting post POWER SUPPLY CONNECTION The power supply connection requires a 12 VDC power supply (included) with at least 300 ma current. The polarity is marked on the instrument.
Introduction Section 1 temperature sensor. All YSI 3200 series cells contain an internal temperature sensor; therefore, an external temperature probe is NOT necessary when using a 3200 series cell. However, if an external temperature probe is used with a YSI 3200 series cell, the external temperature probe will override the temperature sensor in the cell.
SECTION 2 GETTING STARTED 2.1 UNPACKING When you unpack your new YSI Model 3200 for the first time, check the packing list to make sure you have received everything you should have. If there is anything missing or damaged, call the dealer from whom you purchased the Model 3200. If you do not know which of our authorized dealers sold the system to you, call YSI Customer Service at 800-765-4974 or 937767-7241, and we'll be happy to help you. 2.
Getting Started Section 2 Units OPERATION 0.0000 uS K=1.000 25.0°C 11:33AM 1.91%/°C T CELL CELL -10.00°C SETUP DIAG RANGE Soft key functions It is important to remember that the cell constant is NOT used when displaying conductance. To change the measurement units, see 3.5 Display Configuration, Measurement Units. The bottom row of the display shows the soft key menu. 2. Set the Date and Time Press the [SETUP] soft key to enter the Setup menu. SETUP 0.0000 uS/cm K=1.000 25.0°C 11:33AM 1.
Getting Started Section 2 3. Connect the Cell Plug the cell (YSI 3200 series) into the connector on the back of the instrument marked cell. NOTE: If you are using a YSI 3400 series cell, you will need the following: • 3232 adapter • YSI 3220 or 700 series temperature probe 4. Configure the 3200 Next, you must configure the 3200 with the correct cell constant of the cell you are using. From Operation mode, press the [CELL] soft key to enter the Cell menu.
Getting Started Section 2 5. Measurement After setting up the 3200 instrument and cell as described above, the following basic steps should be used to make measurements. 1. Immerse the cell (YSI 3200 series) in the solution to be measured. If using a YSI 3400 series cell, also place the temperature probe in the solution. 2. Gently tap the cell to remove any air bubbles and dip the cell in the solution 2 or 3 times to ensure proper wetting.
SECTION 3 SYSTEM CONFIGURATION AND OPERATION 3.1 MENU The following section shows the software flow chart for the 3200. The two operating modes, Data and Operation, are shown at the top with the soft key functions that are accessed in each mode below them. Data Operation Cell Store Up Configure Down Cal K Digit Offset Store Remove Measure Review Single Point Recall Multi Point Up Send Down Temp Digit L. Temp Send Send All Point 1 Clear Point 2 N. L.
System Configuration and Operation 3.2 Section 3 CELL CALIBRATION Before operating the 3200, or whenever you change cells, you must configure the 3200 to match the cell used. There are two ways to do this. Choose the one that best meets your needs. 1. You may directly enter the manufacturer’s stated (or your manually calculated) cell constant (K). See 3.2 Cell Calibration, Configure. or 2. Allow the 3200 to calculate the cell constant and calibrate the system using standard conductivity solution(s).
System Configuration and Operation Section 3 [0] # Configuration number CONFIG [CONFIG 0] Title [Single] Cell Cal [Linear] Comp [0.68] T.D.S. mult ENTER TO SEND CONFIGURATION UP DOWN 0.0000uS offset [1.00] K [1.91] %/°C Comp val [25.0] °C Comp Temp DIGIT NEXT Press the [UP] or [DOWN] soft key to select one of the six (0-5) stored configurations.
System Configuration and Operation Section 3 Field Range / options Description %/°C comp val 0.0 - 10 Linear temperature compensation value. This value is only used when the Comp field is set to Linear. °C comp temp 0.0 - 100 Reference temperature that readings are compensated to when temperature compensation is used. After you have finished changing the current configuration, press [MODE] two times to return to Operation mode.
System Configuration and Operation Section 3 Entered standard solution CAL CELL K Measured K [961.00] uS/cm Conductivity 1.024 measured K Current temperature 23.0°C [1.00] K Cell ENTER TO STORE UP DOWN DIGIT Multi Point To calculate and store the cell constant (K) at multiple points using different value standard solutions: 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. From the Calibrate K menu, press the [MULTI PT] soft key.
System Configuration and Operation Section 3 OPERATION 3.150 uS/cm Km=1.124 25.0°C 11:33AM 1.50%/°C T CELL CELL 24.38°C SETUP DIAG Current cell constant RANGE Offset When measuring conductance below 10uS (resistance above 100KΩ), the insulation resistance of the cell leads may be a source of error. This error may be greatly reduced by allowing the 3200 to calculate and store a zero offset value. From the Calibrate K menu, press the [OFFSET] soft key to enter the Offset menu.
System Configuration and Operation Section 3 3. Measure a linear temperature compensation factor. 4. Measure a non-linear temperature compensation curve. The previous section, 3.2 Cell Calibration, describes the first two options. The following section, Temp, describes how options 3 and 4 are used. TEMP From the Cell menu, press the [TEMP] soft key to enter the Cell Temp menu.
System Configuration and Operation Section 3 4. Place the cell (and temperature probe, if used) in the solution at the first desired temperature point and wait for the reading to stabilize. Ensure that the cell and temperature probe are submerged and that no air bubbles are trapped in the cell. 5. Press the [POINT 1] soft key to store point 1. 6. Place the cell (and temperature probe, if used) in the same solution at the second desired temperature point and wait for the reading to stabilize. 7.
System Configuration and Operation Section 3 Send From the Cell Temp menu, the [SEND] soft key may be used to send the current temperature compensation data to a computer or serial printer via the RS232 port. 3.4 PLATINIZATION The 3200 can be used to replatinize the electrodes of the cell. See section 6.2 Platinization for details. 3.5 DISPLAY CONFIGURATION The 3200 display is extremely flexible. The user may select from two screen sizes and may select what data is displayed in each field.
System Configuration and Operation 684 Section 3 mg/L TDS 0.4 ppt SAL 1.91%/°C T CELL ENTER TO CHANGE UNITS SIZE NEXT LIGHT DARK MEASUREMENT UNITS Each of the six fields on the small display (three fields on the large display) may be setup by the user.
System Configuration and Operation Section 3 CONTRAST The display contrast is adjusted from the Setup Display menu. SETUP DISP 13.15 uS/cm 11:33AM K=0.100 25.0°C 24.38°C 1.50%/°C T CELL ENTER TO CHANGE UNITS SIZE NEXT LIGHT DARK Press the [LIGHT] or [DARK] soft keys to change the display contrast as necessary. TIME/DATE From Operation mode, press the [SETUP] soft key, then press [TIME/DATE] to enter the Time/Date screen. This screen is used to set the time and date values and format.
System Configuration and Operation Section 3 1. From Operation mode, press the [SETUP] soft key, then, press [SYSTEM] to enter the Setup Sys menu. The following menu will be displayed. [Yes] key beep SETUP SYS [Off] alarm status [0.00] low [0.00] high [Conductance] uS alarm source UP [Off] beep on alarm [On] send on alarm DOWN DIGIT NEXT 2. Use the [NEXT] soft key to select the Alarm status field, then press [UP] or [DOWN] to change the Alarm status to “On”. 3.
System Configuration and Operation Section 3 Since the cursor is already in the Key beep field, press [UP] or [DOWN] to change the status of the beeper. Press [MODE] two times to return to Operation mode. RANGING The Range menu allows the user to select between Auto and Manual ranging. The default is autoranging, where the 3200 will automatically select the range that displays the highest resolution for the sample being measured.
System Configuration and Operation Section 3 OPERATION 13.2uS 11:34AM 24.38°C K=0.100 25.0°C 1.50%/°C T CELL CURRENT RANGE [9.50 - 99.99uS] CELL SETUP DIAGNOSTICS RANGE If the range selected cannot display the current reading, an “OUT OF RANGE” message is displayed as shown below. OPERATION 9.999uS 11:34AM 24.38°C K=0.100 25.0°C 1.
System Configuration and Operation Section 3 conductance resistance conductivity resistivity 1017 uS/cm 984.6 Ωcm mg/L 691 TDS 10.3mS 97.077 Ω ppt 0.5 SAL OPERATION K=0.098 25.0°C 1.91%/°C T CELL CURRENT RANGE [95.0mS - 3.00S] CELL SETUP DIAG RANGE If the [RANGE] soft key is pressed again, the hardware switches to the 0 - 0.9999uS range. The conductance and resistance fields are now out of range and the 3200 displays a warning.
System Configuration and Operation YSI Incorporated Section 3 Model 3200 26
SECTION 4 DATA STORAGE AND RETRIEVAL From Operation mode, press the [MENU] key to enter Data mode. Data mode is used to store, review or delete readings and to send the current reading to a computer or serial printer. 31.15 uS/cm K=1.100 25.0°C 11:33AM 1.50%/°C T CELL STORE 4.1 DATA 24.38°C REVIEW SEND SND SETUP STORE From the data menu, press the [STORE] soft key to enter Store mode and display the following menu. STORE 3.150 uS/cm 11:33AM UP 24.
Data Storage and Retrieval Section 4 If you wish to store a record at a location other than the current one, use the [UP], [DOWN] and [DIGIT] soft keys to select the memory location desired. If the current record location contains data, you will see the following display to warn you that the current memory location contains data. STORE 3.152 uS/cm 11:33AM UP Record # [01] FULL 24.
Data Storage and Retrieval Section 4 RECALL From the Review menu, press the [RECALL] soft key to enter Recall mode and display the following menu. 2/19/97 13.15 uS/cm 69.70 Kohm-cm UP 8:45AM 24.38°C 0.4 ppt DOWN DIGIT RECALL record # [ 0] FULL K=1.090 1.91%/°C SEND The Recall menu is used to recall or send data that was previously stored. The current record number is shown on the right side of the display along with its status.
Data Storage and Retrieval Section 4 NOTE: It is not necessary to clear records before new data can be stored. The new data will overwrite the previous data. 1/ 1/95 0.0 0.0 UP 12:00AM 0.00°C 0.0 DOWN CLEAR record # [ 0] EMPTY K=1.090 1.91%/°C DIGIT DELETE NOTE: There is NO way to restore data once it has been deleted. When you have finished deleting records, press [MODE] three times to return to Operation mode.
Data Storage and Retrieval 4.3 Section 4 SEND While in the Data menu, the [SEND] soft key is used to send the current live display readings to a computer or serial printer via the RS232 port. This is an alternative to storing results, recalling them, and then printing them. DATA 3.150 uS/cm K=10.100 25.0°C 24.38°C 1.91%/°C T CELL STORE 317.5KΩ-cm REVIEW SEND SND SETUP Each time [SEND] is pressed, the current readings are sent to a computer or serial printer via the RS232 port.
Data Storage and Retrieval Section 4 The baud rate range is 1200 to 19200. After you have selected the parameters, press [MODE] once, to return to the Send Setup menu, or three times to return to Operation mode. To select the report format and which data fields will be sent over the RS232 serial port when the [SEND] soft key is pressed, press the [SET SEN] soft key. The following screen will be displayed.
Data Storage and Retrieval Section 4 Comma Delimited Format, send (recalled data) 0 ,3/20/97, 5:37:17PM, 1234uS/cm ,23.45C ,K=0.09,1.91%/C, 9999uS 1 ,3/20/97, 5:46:15PM, 1234uS/cm ,23.45C ,K=0.09,1.91%/C, 9999uS 2 ,3/20/97, 5:57:34PM, 1234uS/cm ,23.45C ,K=0.09,1.91%/C, 9999uS The two fields on the right of the 3200 setup send display, below temperature, allow the user to select the measurement parameters that will be sent.
Data Storage and Retrieval Section 4 To change the S interval, press the [NEXT] soft key to select the S interval field. Then, use the [UP], [DOWN] and [DIGIT] soft keys to enter the S interval parameter.
SECTION 5 DIAGNOSTICS The YSI 3200 has a diagnostic mode that is used to test RS232 communications and the internal beeper. From the Operation menu, press the [DIAG] soft key to enter Diagnostics mode. The following screen will be displayed. DIAG 2.0 Application version 1.4 Controller version SEND BEEP This screen shows the current software version numbers and has soft keys for testing the RS232 communications and the internal beeper.
Diagnostics YSI Incorporated Section 5 Model 3200 36
SECTION 6 MAINTENANCE 6.1 CELL CLEANING AND STORAGE The single most important requirement for accurate and reproducible results in conductivity measurement is a clean cell. A dirty cell will change the conductivity of a solution by contaminating it. To clean a conductivity cell: 1. Dip or fill the cell with cleaning solution and agitate for two to three minutes. Any one of the foaming acid tile cleaners, such as Dow Chemical Bathroom Cleaner, will clean the cell adequately.
Maintenance Section 6 2. Use the [UP], [DOWN] and [DIGIT] soft keys to enter the platinization time required in minutes. Six minutes is usually sufficient for most electrodes. 3. Immerse the cell in the platinizing solution (YSI 3140). Make sure that both electrodes are submerged. 4. Press [ENTER] to start the platinization process. The following screen, showing the time remaining and the polarity of the polarization current, will be displayed. PLATINIZE 5.
SECTION 7 TROUBLESHOOTING 7.1 ERROR MESSAGES The instrument performs a Power On Self Test each time it is turned on. The following error messages are provided to facilitate troubleshooting. They appear on the bottom line of the LCD when an error is detected. TEMP OUT OF COMPENSATION RANGE The unit is configured to use a non-linear temperature compensation curve (either measured or pre-stored). The current measurement temperature is outside the limits of the temperature compensation curve. i.e.
Troubleshooting YSI Incorporated Section 7 Model 3200 40
SECTION 8 PRINCIPLES OF OPERATION 8.1 RATIOMETRIC CONDUCTIVITY The basic concept of microprocessor based ratiometric conductivity measurement is to place an unknown conductance and a known conductance in a series circuit with an AC voltage source, measure the voltage across each conductance and compute the value of the unknown. The object of conductivity measurement is to measure the resistive portion of the complex impedance of the cell.
Principles of Operation Section 8 Looking at our electrodes as sides of a cube, it becomes apparent that the conductance changes as the geometry of the cube changes. If the cube lengthens with respect to the area of the sides, then the conductance will decrease. If the area of the sides increases with respect to the distance between them, then the conductance will increase.
Principles of Operation Section 8 or less have larger electrodes that are closely-spaced. CHOOSING A CONDUCTIVITY CELL Decide which cell will be the most useful for your conditions by considering the conductivity of the solution you want to measure and the size of the sample. 3252, 3418 3256, 3255, 3446 3401 3440 3253, 3254, 0 0.1 1.0 10 100 1000 µS/cm 1 10 100 1000 mS/cm Conductivity The chart above reflects general guidelines. Refer to cell specifications for details.
Principles of Operation Section 8 3252 3253 3254 YSI Incorporated Model 3200 3255 3256 44
Principles of Operation Section 8 CELL CONSTANT CALCULATION YSI 3200 and 3400 Series conductivity cells are calibrated to ± 1% of nominal by means of a YSI transfer standard traceable to OIML Recommendation 56 and NIST. Anytime the condition of the conductivity cell changes, it is possible that the cell constant has also changed. Therefore, you should calibrate your system regularly (See section 3.2 Cell Calibration, Cal K).
Principles of Operation Section 8 CONDUCTIVITY LAW SOLUTION CONDUCTIVITY S/cm or mho/cm mS/cm or mmho/cm µS/cm or µmho/cm CELL CONSTANT = = = = INSTRUMENT CONDUCTANCE S or mho mS or mmho µS or µmho × × × CELL CONSTANT 1/cm 1/cm 1/cm SOLUTION CONDUCTIVITY METER CONDUCTANCE METER CONDUCTANCE = SOLUTION CONDUCTIVITY CELL CONSTANT LOW CONDUCTIVITY MEASUREMENTS When measuring reagent grade water (deionized) or other substances having extremely low conductivity, it is recommend that a flow-through cell
Principles of Operation Section 8 GENERAL CONDUCTIVITY MEASUREMENT PRECAUTIONS After selecting the proper cell, observe the following precautions to ensure accurate, repeatable results: 1. The cell must be clean before making any measurements. When working with substances having low conductivity, extraordinary cleanliness may be required. 2. Soak cells that have been stored dry in deionized water before use. 3. Immerse the cell in the solution deep enough to submerge the vent hole. 4.
Principles of Operation Section 8 temperature coefficient. In extreme cases, the temperature coefficient may have a value as high as 7%/°C. Each conductive ion has a different temperature coefficient. When practical, control the temperature of the solution to be analyzed. For high precision work (±1%), maintain the temperature at 25°C ± 0.1°C. For routine lab work, 25°C ± 0.5°C may be acceptable.
Principles of Operation Section 8 SMALL SAMPLE MEASUREMENTS It is not always possible to immerse the conductivity cell in a solution for measurements. If the quantity of solution is not sufficient for a proper measurement with a dip cell, a sample must be removed for assay. For this application, use the 3254 fill cell. This cell requires 5 ml of sample. Alternatively, any 3200 or 3400 Series cell, except the 3418, may be inverted and used as a fill cell.
Principles of Operation 8.3 Section 8 SALINITY Salinity is determined automatically from the Model 3200 conductivity and temperature readings according to algorithms found in Standard Methods for the Examination of Water and Wastewater (ed. 1995). The use of the Practical Salinity Scale 1978 results in values which are unitless, since the measurements are carried out in reference to the conductivity of standard seawater at 15°C.
Principles of Operation Section 8 ELECTRICAL-NOISE ERRORS Electrical noise can be a problem in any measurement range, but will contribute the most error and be the most difficult to eliminate when using the lowest conductance settings. The noise may be either line-conducted or radiated or both, and may require revised lead dress, grounding, shielding, or all three. Often, all that is necessary is to make sure that parallel leads are of equal length and twisted together.
Principles of Operation YSI Incorporated Section 8 Model 3200 52
SECTION 9 WARRANTY AND REPAIR YSI Model 3200 Instruments are warranted for two years from date of purchase by the end user against defects in materials and workmanship. YSI cells and cables are warranted for one year from date of purchase by the end user against defects in material and workmanship. Within the warranty period, YSI will repair or replace, at its sole discretion, free of charge, any product that YSI determines to be covered by this warranty.
Warranty and Repair Section 9 AUTHORIZED U.S. SERVICE CENTERS FOR THE NEAREST AUTHORIZED SERVICE CENTER, CONTACT: YSI Technical Support • 1725 Brannum Lane • Yellow Springs, Ohio 45387 Phone: +1 (937) 767-7241 • (800) 897-4151 (US) • Fax: (937) 767-1058 Email: environmental@ysi.com • www.ysi.
Warranty and Repair YSI Incorporated Section 9 Model 3200 55
Warranty and Repair Section 9 CLEANING INSTRUCTIONS NOTE: Before they can be serviced, equipment exposed to biological, radioactive, or toxic materials must be cleaned and disinfected. Biological contamination is presumed for any instrument, probe, or other device that has been used with body fluids or tissues, or with wastewater. Radioactive contamination is presumed for any instrument, probe or other device that has been used near any radioactive source.
Warranty and Repair Section 9 PACKING INSTRUCTIONS 1. Clean and decontaminate items to insure the safety of the handler. 2. Complete and include the Cleaning Certificate. 3. Place the product in a plastic bag to keep out dirt and packing material. 4. Use a large carton, preferably the original, and surround the product completely with packing material. 5. Insure for the replacement value of the product.
Warranty and Repair 9.1 Section 9 SOFTWARE UPGRADE PROCEDURE If necessary, the YSI Model 3200’s software can be upgraded using a computer with a serial port. All that is needed is any standard communications software, new 3200 software and a straight RS232 serial cable (not a null modem cable). The 3200 has a female DB-9 connector. A 9 to 25 pin adapter will also be needed if the computer has a 25 pin serial port. The instructions are as follows: 1. 2. 3. 4.
Warranty and Repair 9.2 Section 9 DISASSEMBLY/ASSEMBLY PROCEDURES Place the instrument face down on a padded surface for all procedures below. Case Disassembly • While applying slight separation force to the front, curved edge of the case near one corner, use a small straightblade screwdriver to release the snap (A) on the same side. • When that snap releases, keep applying the separation force, and use the screwdriver to release the front snap (B) nearest the same corner.
Warranty and Repair 9.3 Section 9 TEST AND VERIFICATION PROCEDURE Connect the YSI Model 3200 as shown in the Test and Verification Figure and follow the charts below to verify its accuracy. Equipment Required: If using the YSI 3166 Precision Calibrator Set to simulate conductance, a YSI 3232 adapter is required for proper connection to the cell connector. When using the 3232 adapter, temperature can only be accessed via the EXT. TEMP connector which requires a .25” three conductor phone plug.
Warranty and Repair Section 9 Temperature Verification Temperature Cell Connector Pin 1 & 2 EX. TEMP Connector Displayed Input Resistance ** Input Resistance ** Temperature 0 oC 32660 Ω ± .25% 19600 Ω ± .25% 0.00 ± .15 oC 25 oC 10000 Ω ± .25% 6000 Ω ± .25% 25.00 ± .15 oC 40 oC 5329 Ω ± .25% 3197 Ω ± .25% 40.00 ± .15 oC 70 oC 1752 Ω ± .25% 1051 Ω ± .25% 70.00 ± .20 oC 100 oC 678.5 Ω ± .25% 407.1 Ω ± .25% 100.00 ± .20 oC ** Input temperature resistances either via the EXT.
Warranty and Repair Section 9 Test and Verification Figure Temp. Probe Thermistor 6KΩ @ 25 oC 3200’s External Temperature Connector Decade Resistance Box T RS-232 CELL 12.0 VDC 1 EXT. TEMP 2 3 4 5 6 7 3200 Rear Panel Black 1 Cell Thermistor 10K Ω@ 25 oC Green T 2 Decade Resistance Box Red 3 Brown (Drive) 4 Blue (Sense) YSI 3166 or Decade Resistance Box 5 White (Sense) 6 Yellow (Drive) 7 3200’s 7-Pin Mini DIN Cell Connector YSI Incorporated Cable, Mini DIN Conn.
SECTION 10 ACCESSORIES AND REPLACEMENT PARTS YSI Item # Description 3208 Power Supply, 115 VAC 3209 Power Supply, 240 VAC 003213 Overlay, Window 051016 Overlay, Keypad 051009 Window 111215 Board Assy, PC, Main 111021 Display Assy 111012 Case Assy, Upper 111027 Case Assy, Lower 003226 Weight, SS 051043 Foot, Rubber, Self-Stick 032061 Gasket, Connector, Cell 032062 Gasket, Connector, Ext. Temp. 032063 Gasket, Connector, Power 032064 Gasket, Connector, RS232 051025 Standoff, .
Accessories and Replacement Parts Section 10 10.1 YSI CONDUCTIVITY CELLS YSI 3200 series conductivity cells have a built in temperature sensor for automatic temperature compensation. Dip, fill and flow-through conductivity cells are available, each utilizing platinized platinum iridium electrodes. These cells have the following specifications: Part cgs Cell Number Constant 3200 Series Dip Cells 3252 1.0/cm SI Cell Constant Material Overall Length O.D. Chamber I.D.
Accessories and Replacement Parts Section 10 10.2 STANDARD CALIBRATOR SOLUTIONS YSI manufactures NIST-traceable conductivity calibrator solutions for calibration purposes. The following conductivity calibrator solutions are available from YSI. Part Number 3161 3163 3165 3167 3168 3169 YSI, Incorporated Size 1 quart 1 quart 1 quart 8 pints 8 pints 8 pints Conductivity at 25.00°C 1,000 µ mho/cm ± 0.50% 10,000 µ mho/cm ± 0.25% 100,000 µ mho/cm ± 0.25% 1,000 µ mho/cm ± 1.0% 10,000 µ mho/cm ± 1.
Accessories and Replacement Parts YSI, Incorporated Section 10 Model 3200 66
APPENDIX A - SPECIFICATIONS Modes Conductivity Conductance Resistivity Resistance Salinity TDS Temperature Conductance Range 0 - 0.9999 µS 0.950 - 9.999 µS 9.50 µS - 99.99 µS 95.0 - 999.9 µS 950 - 9999 µS 9.50 - 99.99 mS 95.0 - 999.9 mS 0.95 - 3.00 S Accuracy ± 0.30% full scale ± 0.20% full scale ± 0.10% full scale ± 0.10% full scale ± 0.10% full scale ± 0.10% full scale ± 0.30% full scale ± 1.0% full scale Resolution 0.0001 µS 0.001 µS 0.01 µS 0.1 µS 1 µS 0.01 mS 0.1 mS 0.
Specifications Temperature Compensation Appendix A Method Ref. temp., °C Temp. Coef. Linear, nonlinear 0 - 100 0 - 10%/°C, nonlinear Cell Information Storage 6 Configurations Data Storage 100 points Cell constant, cm-1 0.
APPENDIX B REQUIRED NOTICE This equipment generates and uses radio frequency energy and if not installed and used properly, may cause interference to radio and television reception. There is no guarantee that interference will not occur in a particular installation.
Required Notice YSI, Incorporated Appendix B Model 3200 70
APPENDIX C - RS232 CONNECTIONS Serial Printer Connection To connect to a serial printer, use a 9-pin male to 25 pin male cable. 3200: 9 pin d-sub (male) Printer: 25 pin d-sub (male) pin 2 pin 2 pin 5 pin 7 NOTE: Only two connections are required for a serial printer. The following printer, manufactured by Nippon Primex Co., Ltd.
RS232 Connections YSI, Incorporated Appendix C Model 3200 72
APPENDIX D TEMPERATURE CORRECTION DATA FOR TYPICAL SOLUTIONS A. Potassium Chloride** (KCl) Concentration: 1 x 10-1 mole/liter Concentration: 1 mole/liter C 0 5 10 15 20 25 mS/cm %/ C (to 25 C) 65.10 73.89 82.97 92.33 101.97 111.90 1.67 1.70 1.72 1.75 1.77 1.80 Concentration: 1 x 10-2 mole/liter C mS/cm %/ C (to 25 C) 0 5 10 15 20 25 30 35 37.5 40 45 50 7.13 8.22 9.34 10.48 11.65 12.86 14.10 15.38 16.04 16.70 18.05 19.43 1.78 1.80 1.83 1.85 1.88 1.90 1.93 1.96 1.98 1.99 2.02 2.
Temperature Correction Data for Typical Solutions Appendix D B. Sodium Chloride* (NaCl) Saturated solutions at all temperatures C 0 5 10 15 20 25 30 mS/cm %/ C (to 25 C) 134.50 155.55 177.90 201.40 225.92 251.30 277.40 1.86 1.91 1.95 1.99 2.02 2.05 2.08 Concentration: 1 x 10-1 mole/liter Concentration: 0.5 mole/liter C mS/cm %/ C (to 25 C) 0 5 10 15 20 25 30 35 37.5 40 45 50 25.90 29.64 33.61 37.79 42.14 46.65 51.28 56.01 58.40 60.81 65.65 70.50 1.78 1.82 1.86 1.90 1.93 1.96 1.99 2.01 2.02 2.
Temperature Correction Data for Typical Solutions Appendix D C. Lithium Chloride* (LiCl) Concentration: 1 x 10-1 mole/liter Concentration: 1 mole/liter C mS/cm %/ C (to 25 C) 0 5 10 15 20 25 30 35 37.5 40 45 50 39.85 46.01 52.42 59.07 65.97 73.10 80.47 88.08 91.97 95.92 103.99 112.30 1.82 1.85 1.89 1.92 1.95 1.98 2.02 2.05 2.07 2.08 2.11 2.15 Concentration: 1 x 10-2 mole/liter C mS/cm %/ C (to 25 C) 0 5 10 15 20 25 30 35 37.5 40 45 50 5.07 5.98 6.87 7.75 8.62 9.50 10.40 11.31 11.78 12.26 13.
Temperature Correction Data for Typical Solutions Appendix D E. Ammonium Chloride* (NH4Cl) Concentration: 1 x 10-1 mole/liter Concentration: 1 mole/liter C 0 5 10 15 20 25 mS/cm %/ C (to 25 C) 64.10 74.36 83.77 92.35 100.10 107.00 1.60 1.53 1.45 1.37 1.29 1.21 Concentration: 1 x 10-2 mole/liter C mS/cm %/ C (to 25 C) 0 5 10 15 20 25 30 35 37.5 40 45 50 6.96 7.98 9.09 10.27 11.50 12.78 14.09 15.43 16.10 16.78 18.12 19.45 1.82 1.88 1.93 1.97 2.00 2.03 2.06 2.07 2.08 2.08 2.09 2.
APPENDIX E - CONVERSION FACTORS TO CONVERT FROM TO EQUATION mhos Siemens Multiply by 1 mhos ohms 1/mho ohms mhos 1/ohm Feet Meters Multiply by 0.3048 Meters Feet Multiply by 3.
Conversion Factors YSI, Incorporated Appendix E Model 3200 78
APPENDIX F - GLOSSARY OF TERMS ampere (A) - SI unit of electric current; one coulomb per second. amplitude - The maximum deviation of an alternating current from its average value during its cycle. ASTM - American Society for Testing and Materials calibrate - To determine, check, or rectify the graduation of any instrument giving quantitative measurements. calibrator solution - A solution of known value used to calibrate.
Glossary of Terms Appendix F OIML - Organisation Internationale de Métrologie Légale is a treaty organization for the harmonization of practical applications of measurement standards. platinum black - Platinum precipitated from a solution of the (IV) chloride by reducing agents. A velvety-black powder. polarization - The separation of the positive and negative charges of a molecule by an external agent. Pyrex - A trademark name for heat-resistant and chemical-resistant glass.
Glossary of Terms YSI Incorporated Appendix F Model 3200 81
1700/1725 Brannum Lane Yellow Springs, Ohio 45387 USA (800) 765-4974 • (937) 767-7241 FAX (937) 767-9353 Email: environmental@ysi.com Website: www.YSI.