METEOROLOGICAL INSTRUMENTS INSTRUCTIONS TEMP TRACKER MODEL 46203 R.M. YOUNG COMPANY 2801 AERO PARK DRIVE, TRAVERSE CITY, MICHIGAN 49686, USA TEL: (231) 946-3980 FAX: (231) 946-4772 WEB: www.youngusa.
MODEL 46203 TEMP TRACKER FRONT PANEL INTRODUCTION The YOUNG Model 46203 Temp Tracker is a compact temperature and humidity display with advanced alarm features. Its features include: • • • • • • • • 1. Temperature display 2. Temperature units 3. Min / Max indicators 4. RH or Dew Point display 5. Alarm status indicators 6. Brightness control (enter key in SETUP) 7.
MOUNTING AND STARTUP 1. Select location for display. A location that is out of direct sunlight provides best visibility. 2. Connect cables to terminals. Refer to wiring diagram. 3. Connect GND terminal to suitable earth ground. 4. Insert power supply plug into power jack, plug AC adapter into wall outlet. 5. The Temp Tracker will display a software version number for several seconds. It will then begin to display temperature & humidity information.
MINIMUM / MAXIMUM Scale Temperature and Dew Point: -50 to 50 °C, -50 to 150 °F Relative Humidity: 0 to 100% • Alarm relays Normally open, non-latching, 5A resistive, 2A inductive @ 24 VAC or 30 VDC The Temp Tracker may be set to automatically cycle through MIN, MAX and Current values. This feature is selected during SETUP. • Power requirement 12 to 30 VDC, 4.5 watts Minimum and maximum values reset if power is interrupted. • Weight 0.75 lbs (0.
46203-90 Page 4