Manual

TYPICAL WIRING
+PWR
GND
RH
T
(WHT)
(BLK)
(GRN)
(RED)
CURRENT INPUT 1 (+)
CURRENT INPUT 2 (+)
+24VDC PWR
PWR COMMON
SHIELDED CABLE
EARTH GROUND
4-20 mA CURRENT INPUT
MEASURING INSTRUMENT
SENSOR TERMINALS
CURRENT INPUT COMMON
250 Ohms Typical
R R
+PWR
GND
RH
T
(WHT)
(BLK)
(GRN)
(RED)
VOLTAGE INPUT 1 (+)
VOLTAGE INPUT 2 (+)
+24VDC PWR
PWR COMMON
SHIELDED CABLE
EARTH GROUND
VOLTAGE INPUT
MEASURING INSTRUMENT
VOLTAGE INPUT COMMON
50 to 250 Ohms
Typical
R R
SENSOR TERMINALS
30VDC at 46 mA Max Supply
Total Current Consumption = T mA + RH mA + 6 mA
Min Supply V = 4 + 0.02R
Max Load R = (Vsupply - 4) / 0.02