Manual
  <7.GeneralSpecications>
7-3
IM 01C50B01-01E
Table 7.1  Input Type, Measurement Range and Accuracy
Sensor Type
Reference
Standard
Measurement Range
Minimum 
Span
(Recommended)
Accuracy
Input range A/D Accuracy
D/A
Accuracy
°C °F °C °F °C °F
T/C
B
IEC584
100 to 1820 212 to 3308
25°C
(45°F)
100
300
400
to
to
to
300
400
1820
212 
572
752
to
to
to
572
752
3308
±3.0
±1.0
±0.75
±5.4
±1.8
±1.35
±0.02%
of span
E -200 to 1000 -328 to 1832
-200
-50
to
to
 -50
1000
-328
-58
to
to
-58
1832
±0.35
±0.16
±0.63
±0.29
J -200 to 1200 -328 to 2192
-200
-50
to
to
50
1200
-328
-58
to
to
-58
2192
±0.40
±0.20
±0.72
±0.36
K -200 to 1372 -328 to 2502
-200
-50
to
to
-50
1372
-328
-58
to
to
-58
2502
±0.50
±0.25
±0.90
±0.45
N -200 to 1300 -328 to 2372
-200
-50
to
to
-50
1300
-328
-58
to
to
-58
2372
±0.80
±0.35
±1.44
±0.63
R -50 to 1768 -58 to 3214
-50
0
100
600
to
to
to
to
0
100
600
1768
-58
32
212
1112
to
to
to
to
32
212
1112
3214
±1.0
±0.80
±0.60
±0.40
±1.8
±1.44
±1.08
±0.72
S -50 to 1768 -58 to 3214
-50
0
100
600
to
to
to
to
0
100
600
1768
-58
32
212
1112
to
to
to
to
32
212
1112
3214
±1.0
±0.80
±0.60
±0.40
±1.8
±1.44
±1.08
±0.72
T -200 to 400 -328 to 752
-200
-50
to
to
-50
400
-328
-58
to
to
-58
752
±0.25
±0.14
±0.45
±0.25
W3
ASTM
E988
0 to 2300 32 to 4172
0
400
1400
2000
to
to
to
to
400
1400
2000
2300
32
752
2552
3632
to
to
to
to
752
2552
3632
4172
±0.80
±0.50
±0.60
±0.90
±1.44
±0.90
±1.08
±1.62
W5 0 to 2300 32 to 4172
0
400
1400
2000
to
to
to
to
400
1400
2000
2300
32
752
2552
3632
to
to
to
to
752
2552
3632
4172
±0.70
±0.50
±0.70
±0.90
±1.26
±0.90
±1.26
±1.62
L
DIN43710
-200 to 900 -328 to 1652
-200
-50
to
to
-50
900
-328
-58
to
to
-58
1652
±0.30
±0.20
±0.54
±0.36
U -200 to 600 -328 to 1112
-200
-50
to
to
-50
600
-328
-58
to
to
-58
1112
±0.50
±0.25
±0.90
±0.45
RTD
Pt100
Pt200
Pt500
JPt100
IEC751
-200
-200
-200
-200
to
to
to
to
850
850
850
500
-328
-328
-328
-328
to
to
to
to
1562
1562
1562
932
10°C
(18°F)
-200
-200
-200
-200
to
to
to
to
850
850
850
500
-328
-328
-328
-328
to
to
to
to
1562
1562
1562
932
±0.14
±0.30
±0.20
±0.16
±0.25
±0.54
±0.36
±0.29
JIS C1604
Cu
SAMA
RC21-4
-70 to 150 -94 to 302
-70
-40
to
to
-40
150
-94
-40
to
to
-40
302
±1.35
±1.0
±2.43
±1.8
Ni120 — -70 to 320 -94 to 608 -70 to 320 -94 to 608 ±0.11 ±0.19
mV
—
-10 to 100 [mV] 3 [mV]
—
±12 [μV]
ohm 0 to 2000 [Ω] 20 [Ω] ±0.35 [Ω]
Table 7.2  YTA110 Effect of Ambient Temperature
Sensor Type TemperatureCoefcient
Thermocouples E, J, K, N, T, L, U 0.08°C + 0.02% of abs.reading
Thermocouples R, S, W3, W5 0.25°C + 0.02% of abs.reading
T/C B
100°C ≤ Reading < 300°C 1°C + 0.02% of abs.reading
300°C ≤ Reading 0.5°C + 0.02% of abs.reading
RTD 0.08°C + 0.02% of abs.reading
mV 0.002 mV + 0.02% of abs.reading
ohm 0.1Ω + 0.02% of abs.reading
N
o
te1:  Ambient Temperature Effect per 10°C change is 
±0.1% or ±(temperature coefcient/span), whichever 
is greater.
No
te2:  The “abs.reading” on Table7.2 means the absolute 
value of the reading in °C. 
   Example of abs reading;
  When the temperature value is 250 Kelvin, abs reading 
is 23.15, absolute (250–273.15).
Example of Ambient Temperature Effect;
   Conditions;
  1) Input Sensor: Pt100 
  2) Calibration Range: –100 to 100°C
  3) Reading value: –50°C
   Ambient Temperature Effect per 10°C;
  Temperature Coefcient/Span
  =(0.08°C+0.02/100×|–50°C|)/{100°C–(–100°C)} 
= 0.00045→ 0.045%
  Therefore, Ambient Temperature Effect is ±0.1%/10°C