Manual
    <6.  Maintenance>
6-1
IM 01C50B01-01E
6.  Maintenance
6.1  General
Each component of this instrument is congured in 
units to make maintenance easier.
This chapter contains disassembly and assembly 
procedures associated with calibration, adjustment 
and part replacement required for maintenance of 
the affected instrument.
IMPORTANT
1. Maintenance of this instrument should be 
performed in a service shop where the 
necessary tools are provided.
2. Handling the CPU assembly  
Some of the parts contained in the CPU 
assembly are susceptible to static electricity 
damage. Before performing maintenance, 
use a ground wrist band or other antistatic 
mea sures, and avoid touching the electronic 
components and circuits with bare hands. 
When removed from the instrument, keep 
the CPU assembly in an antistatic bag.
6.2  Calibration
This instrument is fully factory-tested and is 
guaranteed for the intended accuracy, eliminating 
the need for calibration. When calibration needs to 
be varied, the following equipment and calibration 
procedure is recommended.
6.2.1  Selection of Equipment for 
Calibration
Table 6.1 lists the equipment required for 
calibration. The calibration equipment traceable to a 
verifying agency standard should be used.
6.2.2  Calibration Procedure
To conduct calibration required to evaluate the 
uncertainty while using the instrument, follow the 
steps below:
1.  In accordance with the example wiring shown in 
Figure 6.1, connect each equipment and initiate 
warm up. Lay wiring on the input side according 
to the sensor to be used.
Table 6.1  Calibration Equipment List
Name Recommended Remark
Power 
supply
SDBT, SDBS type 
distributor
4 to 20mA DC
(Output voltage: 
26.5±1.5V, drop 
by internal 250Ω 
resistance included)
Load 
resistance
2792 type standard 
resistor
(250Ω ±0.005%)
For 4 to 20mA DC
Voltmeter Model 1271 
digital multimeter 
(accuracy: ±0.002%)
For 4 to 20mA DC 
signal
Universal 
calibrator
Model 9100 type For calibration of 
DC voltage and 
thermocouple
Variable 
resistor
279301 type 6-dial 
variable resistor 
(accuracy: ±0.001% 
±2mΩ)
For calibration 
of thermometer 
resistor (RTD) input
F0601.ai
1
2
3
4
5
1
2
3
4
5
(A)
(B)
(B)
(A)
a.  Wiring of power supply and output
b. Example of wiring for thermocouple or DC voltage input
  (when 1 input type is used)
+  Output signal
– Load
resistance
DC voltage generator
Voltmeter
c.  Example of wiring for thermometer resistor 4-core type
  (when 1 input type is used)
Variable resistor
(+)
(–)
Figure 6.1  Example of Wiring for Calibration 
Equipment
2.  For DC voltage input
With a voltage generator, deliver input signals 
corresponding to 0, 25, 75, or 100% of the input 
span to the temperature transmitter. Measure 
the resulting input signal with the voltmeter 
(digital multimeter) and check the output value 
relative to the input value.










