Instruction Manual
<9. Parameter Summary>
9-1
IM 01C50E01-01EN
9. Parameter Summary
Table 9.1 Parameter Summary
Object
ID
Attribute
ID
Label Description Default value Handling
1.
UAPMO
block
1 Version Revision Indicates the application revision of YTA. This revision changes
when the application software is downloaded.
--- R
10 Static Revision Indicates the revision level of the xed parameters of UAP.
Used, for example, to check whether parameters have been
changed.
0 R
64 Identication
Number
Indicates the vendor ID, model ID, and revision of the device. --- R
65 CTS Version Indicates the version of the communication stack test system
(CTS).
0 R
66 ITS Version Indicates the version of the interoperability test system (ITS). 0 R
67 Diagnostic Status Indicates the diagnostic results of the device based on the
NAMUR NE107
*1
model.
Setting Enable diagnostic status conguration in UAP Option to
Enable allows turning OFF and ON the display of the diagnostic
results for each summary, and changing Categorize. For
Categorize at the time of shipment, refer to tables 8.3 and 10.2.
--- R
68 UAP Option Allows setting the Diagnostic Status and write protection of UAP.
Software write protect
1: On, 0: Off
Enable hardware write protect
1: Enable, 0: Disable (default)
Enable diagnostic status conguration
1: Enable, 0: Disable (default)
The following table shows the relationship between the
hardware write protection and software write protection.
Enable
hardware
write
protect
Hardware
write
protect
Software write
protect
Write
protect
Disable Off or On Off No
Disable Off or On On Protected
Enable Off Off or On No
Enable On Off or On Protected
0 W
69 Diagnostic Switch Allows setting On/Off for each summary of Diagnostic Status
when Enable diagnostic status conguration in UAP Option is
set to Enable.
On W
70 Diagnostic
Conguration
Allows Categorize for each summary of Diagnostic Status when
Enable diagnostic status conguration in UAP Option is set to
Enable.
Refer to Table
9.2.
W
102 Diagnostic Status
Detail
Detailed information on Diagnostic Status. For Categorize at the
time of shipment, refer to Table 9.2.
Refer to Table
9.2.
R
103 Energy Left Indicates the number of days of remaining battery life assuming
ambient temperature condition as 23 degrees Celsius. The unit
is day.
--- R
104 Reset Energy Left Resets the remaining battery power calculation to restore it to
a remaining battery power calculation which is based on new
batteries.
0 = Continue
1 = Reset
0 (reading
value is
always 0)
W
105 Power Supply
Status
Indicates remaining battery life and power supply of device
0 – line powered
1 – battery powered, greater than 75% remaining capacity
2 – battery powered, between 25% and 75% remaining
capacity
3 – battery powered, less than 25% remaining capacity
--- R
Note: TSAP ID is 2.