User Manual

4-8
IM 04L21B01-17EN
Measured and Computed Data
The FD command is used to output the data.
You can use the CB command to specify whether to output the data of measurement
channels set to skip and computation channels set to OFF.
Syntax
The measured/computed data is output in the following syntax along with the date and
time information for each channel.
EACRLF
DATE_yy/mo/ddCRLF
TIME_hh:mm:ss.mmmtCRLF
s_ccca
1
a
2
a
3
a
4
uuuuuufdddddE-ppCRLF
····························
ENCRLF
yy
Year (
00
to
99
)
mo
Month (
01
to
12
)
dd
Day (
01
to
31
)
hh
Hour (
00
to
23
)
mm
Minute (
00
to
59
)
ss
Second (
00
to
59
)
mmm
Millisecond (
000
to
999
. A period is placed between seconds and
milliseconds.)
t
Reserved (Space.)
s
Data status (
N
,
D
,
S
,
O
,
E
, or
B
)
N:
Normal
D:
Differential input
S:
Skip
O:
Over
E:
Error
B:
Burnout
ccc
Channel number (3 digits)
001
to
012:
Measurement channel
101
to
124:
Computation channel
a
1
a
2
a
3
a
4
a
1
Alarm status (level 1)
a
2
Alarm status (level 2)
a
3
Alarm status (level 3)
a
4
Alarm status (level 4)
(Each status is set to
H
,
L
,
h
,
l
,
R
,
r
,
T
,
t
, or space.)
((
H
: high limit alarm,
L
: low limit alarm,
h
: difference high-limit alarm,
l
:
difference low-limit alarm,
R
: high limit on rate-of-change alarm,
r
: low
limit on rate-of-change alarm,
T
: delay high limit alarm,
t
: delay low limit
alarm, space: no alarm)
uuuuuu
Unit information (6 characters, left-justified)
mV____:
mV
V_____:
V
^C____:
°C
xxxxxx:
(User-defined character string)
f
Sign (
+
,
)
4.2 Output Format of Text Data