User Manual
4-7
IM 04L21B01-17EN
Responses
1
2
3
4
5
6
App
Index
• Example
EA
SR001,VOLT,20mV,0,20
SR002,VOLT,20mV,0,20
···················
EN
Decimal Place/Unit Information
• The FE command is used to output the data.
• You can use the CB command to specify whether to output the data of measurement
channels set to skip and computation channels set to OFF.
•
Syntax
The data is output for each channel in the following syntax.
EACRLF
s_cccuuuuuu,ppCRLF
··················
ENCRLF
s
Data status (
N
,
D
, or
S
)
N:
Normal
D:
Differential input
S:
Skip (When the measurement range is set to SKIP for a
measurement channel or when the channel is turned OFF for a
computation channel)
ccc
Channel number (3 digits)
001
to
012:
Measurement channel
101
to
124:
Computation channel
uuuuuu
Unit information (6 characters, left-justified)
mV____:
mV
V_____:
V
^C____:
°C
xxxxxx:
(User-defined character string)
pp
Decimal place (
00
to
04
)
No decimal (
00000
) for
00
.
One digit to the right of the decimal (
0000.0
) for
01
.
Two digits to the right of the decimal (
000.00
) for
02
.
Three digits to the right of the decimal (
00.000
) for
03
.
Four digits to the right of the decimal (
0.0000
) for
04
.
When you are using a channel that is set to Log scale (/LG1 option), the
decimal place of the FX digital display’s mantissa is displayed.
_
Space
•
Example
EA
N001mV,01
N002mV,01
EN