User Manual

4-7
IM 04L21B01-17EN
Responses
1
2
3
4
5
6
App
Index
Example
EA
SR001,VOLT,20mV,0,20
SR002,VOLT,20mV,0,20
···················
EN
Decimal Place/Unit Information
The FE command is used to output the data.
You can use the CB command to specify whether to output the data of measurement
channels set to skip and computation channels set to OFF.
Syntax
The data is output for each channel in the following syntax.
EACRLF
s_cccuuuuuu,ppCRLF
··················
ENCRLF
s
Data status (
N
,
D
, or
S
)
N:
Normal
D:
Differential input
S:
Skip (When the measurement range is set to SKIP for a
measurement channel or when the channel is turned OFF for a
computation channel)
ccc
Channel number (3 digits)
001
to
012:
Measurement channel
101
to
124:
Computation channel
uuuuuu
Unit information (6 characters, left-justified)
mV____:
mV
V_____:
V
^C____:
°C
xxxxxx:
(User-defined character string)
pp
Decimal place (
00
to
04
)
No decimal (
00000
) for
00
.
One digit to the right of the decimal (
0000.0
) for
01
.
Two digits to the right of the decimal (
000.00
) for
02
.
Three digits to the right of the decimal (
00.000
) for
03
.
Four digits to the right of the decimal (
0.0000
) for
04
.
When you are using a channel that is set to Log scale (/LG1 option), the
decimal place of the FX digital display’s mantissa is displayed.
_
Space
Example
EA
N001mV,01
N002mV,01
EN
4.2 Output Format of Text Data