User Manual

3-31
IM 04L21B01-17EN
Commands
1
2
3
4
5
6
App
Index
EV Executes manual sample,
generates a manual trigger, takes
a snapshot, or causes a timeout
Syntax
EVp1<terminator>
p1
Action type
0
Executes manual sampling
1
Generates a manual trigger
2
Takes a snapshot
3
Causes a timeout in display data
4
Causes a timeout in event data
Example Execute manual sampling.
EV0
Description EV1 is only valid when the key trigger is set
to ON using the event data sample condition
command (the TE command). The EV1
command is equivalent to a key trigger.
CL Executes manual SNTP (/C7
option)
Syntax
CLp1<terminator>
p1
Executes manual SNTP (0)
Example Synchronize the clock.
CL0
CV Switches between the normal
and secondary trend intervals
Syntax
CVp1<terminator>
p1
Trend interval (0, 1)
0
Switches to the first trend interval
(the normal trend interval)
1
Switches to the secondary trend
interval
Example Set the trend interval to the secondary trend
interval.
CV1
MS Writes a message (display and
write)
Syntax
MSp1,p2,p3<terminator>
p1
Message number (1 to 100)
p2
Message write destination
GROUP
A specific display group
ALL
All display groups
p3
Display group number
The display group number when p2 is set
to GROUP.
This parameter has no meaning when p2 is
set to ALL.
Example Write the message in message number 8 to
display group 1.
MS8,GROUP,1
Description If you omit p2, the message is written to all
display groups.
Set p3 by referring to the table in section 3.3.
BJ Writes a free message
Syntax
BJp1,p2,p3,p4<terminator>
p1
Message number (1 to 10)
p2
Message (up to 32 characters)
p3
Message write destination
GROUP
A specific display group
ALL
All display groups
p4
Display group number
The display group number when p2 is set
to GROUP.
This parameter has no meaning when p2 is
set to ALL.
Example
Using message number 3, write the string
ALARM
” to all groups.
BJ3,ALARM,ALL
Description If you omit p3, the message is written to all
display groups.
Set p3 and p4 by referring to the table in
section 3.3.
EJ Changes the password of the
login function (/C7 option)
Syntax
EJp1,p2,p3<terminator>
p1
Old password (up to 8 characters)
p2
New password (up to 8 characters)
p3
New password (up to 8 characters)
Example Change the password from “PASS001” to
“WORD005.”
EJPASS001,WORD005,WORD005
TL Starts, stops, resets
computation (MATH) or clears
the computation dropout status
display (/M1, /PM1, and /PWR1
options)
Syntax
TLp1<terminator>
p1
Action type
0
Starts computation
1
Stops computation
2
Resets computation
3
Clears the computation data
dropout display
Example Start computation.
TL0
Description You cannot use this command while the FX is
saving or loading setup data.
DS Switches between execution
modes (operation and basic
setting)
Syntax
DSp1<terminator>
p1
Mode
0
Operation mode
1
Basic setting mode
3.5 Control Commands