Setup Items
3
2
1
4
5
6
7
8
9
10
11
12
13
14
App
Index
Communication (Ethernet) > FTP Client > FTP connection
Setup Item Selectable Range or Selections Default Value Setting
FTP connection Primary/Secondary Primary
FTP server name 64 characters or less
Port number Numerical value (0 to 65535) 21
Login name 32 characters or less
Password 32 characters or less ******. . . ***
Account 32 characters or less
PASV mode On/Off Off
Initial path 64 characters or less
Communication (Ethernet) > SNTP client
Setup Item Selectable Range or Selections Default Value Setting
Use/Not Use/Not Not
Server name 64 characters or less
Port number Numerical value (0 to 65535) 123
Access interval Off/1h/8h/12h/24h 8h
Access reference time Numerical value (00:00 to 23:59) 00:00
Access timeout 10s/30s/90s 30s
Time adjust on Start action On/Off Off
Communication (Ethernet) > Modbus client > Basic settings
Setup Item Selectable Range or Selections Default Value Setting
Read cycle 1s/2s/5s/10s 1s
Retry interval Off/10s/20s/30s/1min/2min/5min/10min/20min/
30min/1h
2min
Communication (Ethernet) > Modbus client > Modbus server settings
Setup Item Selectable Range or Selections Default Value Setting
Server number 1-8/9-16 1-8
Port Numerical value (0 to 65535) 502
Modbus server name 64 characters or less
Unit Auto/Fixed Auto
No. 0 to 255 1
Communication (Ethernet) > Modbus client > Command settings
Setup Item Selectable Range or Selections Default Value Setting
Client command number 1-8/9-16 1-8
1 to 16 Off/R-M/W/W-M Off
First Depends on the command type
Last Depends on the command type
Server 1/2/3/. . . /15/16 1
Regi. Numerical value (register number) 30001
Type INT16/UINT16/INT32_B/INT32_L/UINT32_B/ INT16
UINT32_L/FLOAT_B/FLOAT_L