14-4
IM 04L21B01-01EN
Setup Items in Setting Mode
Menu tab
Date/Time > Date&Time
Setup Item Selectable Range or Selections Default Value Setting
Date&Time > Time set 2011/01/01
00:00:00
Date/Time > Daylight Saving Time
Setup Item Selectable Range or Selections Default Value Setting
Use/Not Use/Not Not
Start time > Month JAN/FEB/MAR/APR/MAY/JUN/JUL/AUG/SEP/OCT/NOV/DEC MAR
Start time > Day order 1st/2nd/3rd/4th/Last 2nd
Start time > Day of the week SUN/MON/TUE/WED/THU/FRI/SAT SUN
Start time > Hour of the day Numerical value (0 to 23) 2
End time > Month JAN/FEB/MAR/APR/MAY/JUN/JUL/AUG/SEP/OCT/NOV/DEC NOV
End time > Day order 1st/2nd/3rd/4th/Last 1st
End time > Day of the week SUN/MON/TUE/WED/THU/FRI/SAT SUN
End time > Hour of the day Numerical value (0 to 23) 1
Meas channel > Range, Alarm
Setup Item Selectable Range or Selections Default Value Setting
First-CH, Last-CH Channel number 001
Range > Mode Skip/Volt/TC/RTD/Scale/Delta/DI/ Volt
1-5V/LogType1/LogType2/Sqrt
Mode = Volt
Range 20mV/60mV/200mV/2V/6V/20V/50V/1V 2V
Span Lower Numerical value (depends on the range)
Span Upper Numerical value (depends on the range)
Mode = TC
Range R/S/B/K/E/J/T/N/W/L/U/WRe (Options may add additional
settings.)
R
Span Lower Numerical value (depends on the range)
Span Upper Numerical value (depends on the range)
Mode = RTD
Range Pt/JPt (Options may add additional settings.) Pt
Span Lower Numerical value (depends on the range)
Span Upper Numerical value (depends on the range)
Mode = Scale
Type Volt/TC/RTD/DI Volt
Range Depends on the type
Span Lower Numerical value (depends on the range)
Span Upper Numerical value (depends on the range)
Scale Lower Numerical value (–30000 to 30000; decimal place: 0 to 4) 0.00
Scale Upper Numerical value (–30000 to 30000) 200.00
Unit 6 characters or less
Mode = Delta
Type Volt/TC/RTD/DI Volt
Range Depends on the type
Span Lower Numerical value (depends on the range)
Span Upper Numerical value (depends on the range)
Ref.CH Meas channel number
14.1 Setting Mode Menu Map and Setup Items