Manual

13-19
IM 04L21B01-01EN
Specifications
3
2
1
4
5
6
7
8
9
10
11
12
13
14
App
Index
Standard Performance
Item Specifications
Measurement/display accuracy
Standard operating conditions:
Temperature: 23 ± 2°C
Humidity: 55 ± 10%RH
Power supply voltage: 90 to 132 or 180 to 250 VAC
Power supply frequency: 50/60 Hz ± 1%
Warm-up time: At least 30 minutes.
Other ambient conditions such as vibration should not adversely affect the operation.
Input Type Range Measurement Accuracy (Digital display) Digital Display
Max. Resolution
DC voltage
20 mV ±(0.05% of rdg + 12 digits) 1 μV
60 mV
±(0.05% of rdg + 3 digits)
10 μV
200 mV 10 μV
1 V 100 μV
2 V ±(0.05% of rdg + 12 digits) 100 μV
1 to 5 V
±(0.05% of rdg + 3 digits)
1 mV
6 V 1 mV
20 V 1 mV
50 V 10 mV
Thermocouple
(Not including
the accuracy of
reference junction
compensation; when
the burnout detection
function is off.)
R ±(0.15% of rdg + 1°C)
R, S 0 to 100°C: ±3.7°C,
100 to 300°C: ±1.5°C
B 400 to 600°C: ±2°C; accuracy not guaranteed for
temperatures less than 400°C
0.1°C
S
B
K ±(0.15% of rdg + 0.7°C)
–200 to –100°C: ±(0.15% of rdg + 1°C)
E ±(0.15% of rdg + 0.5°C)
–200 to –100°C: ±(0.15% of rdg + 0.7°C)
J
T
N ±(0.15% of rdg + 0.7°C)
–200 to 0°C: ±(0.35% of rdg + 0.7°C)
Accuracy not guaranteed for temperatures less than –200°C.
W ±(0.15% of rdg + 1°C)
L ±(0.15% of rdg + 0.5°C)
–200 to –100°C: ±(0.15% of rdg + 0.7°C)
U
WRe ±(0.2% of rdg + 2.5°C)
0 to 200°C: ±4.0°C
RTD Pt100 ±(0.15% of rdg + 0.3°C)
JPt100
DI Voltage Threshold level (Vth=2.4 V) accuracy ± 0.1 V
Contact With parallel capacitance of 0.01 μF or less, 1 kΩ or less: 1 (ON). 100 kΩ or more: 0 (OFF)
Measuring accuracy in case of scaling
Accuracy during scaling (digits) = measurement accuracy (digits) × multiplier + 2 digits (rounded up)
* Fractions rounded up
where the multiplier = scaling span (digits)/measuring span (digits).
Example For 1-5 V range (A/D integration time is 16.7 ms or more), measurement span of 1.000 to
5.000 V, and scaling span of 0.000 to 2.000
The measuring accuracy for 5 V input is as follows.
Measuring accuracy (1-5 Vrange) = ±(0.05% × 5 V + 3 digits) = ±(0.0025 V [3 digits] + 3
digits) = ±6 digits
Multiplier = {2000 digits (0.000 to 2.000)}/4000 digits (1.000 to 5.000) = 0.5
Thus, accuracy during scaling = ±(6 × 0.5 + 2) digits = 5 digits (rounded up)
Reference junction compensation accuracy
When measuring temperature greater than or equal to 0 °C and when input terminal temperature is
balanced
Type R, S, W, WRe: ±1.0°C
Type K, J, E, T, N, L, and U: ±0.5°C
Type B: Internal reference compensation is fixed to 0°C