Index
13.6 General Specifications
Construction
Item Specifications
Mounting Flush panel mounting (on a vertical plane)
Mounting angle Inclined backward up to 30 degrees from a horizontal plane.
Allowable panel thickness 2 to 26 mm
Material Case: Metal plate
Bezel and display cover: Polycarbonate
Color Case: Grayish blue green (Munsell 2.0B5.0/1.7 or equivalent)
Bezel: Charcoal gray light (Munsell 10B3.6/0.3 or equivalent)
Front panel Water and dust proof: Complies with IEC529-IP65, except side-by-side mounting
External dimensions 144 (W) × 144 (H) × 161.7 (D) mm (D: depth from the panel mounting plane)
Weight FX1002, FX1004, FX1006: 1.3 kg
FX1008, FX1010, FX1012: 1.4 kg
(Not including options)
Normal Operating Conditions
Item Specifications
Supply voltage 90 to 132, 180 to 250 VAC
Power supply frequency 50 Hz ± 2%, 60 Hz ± 2%
Ambient temperature 0 to 50°C
Ambient humidity 20 to 80%RH (at 5 to 40°C), 10 to 50% (at 40 to 50°C)
Vibration 10 to 60 Hz, 0.2 m/s
2
Shock Not allowed
Magnetic field 400 A/m or less (DC and 50/60 Hz)
Noise Normal mode (50/60 Hz)
DC voltage
The peak value including the signal must be less than 1.2 times the measuring range.
Thermocouple
The peak value including the signal must be less than 1.2 times the measuring thermal electromotive force.
RTD 50 mV or less
Common mode noise 250 VACrms or less for all ranges (50/60 Hz)
Maximum noise voltage between channels
FX1xxx-x-x-H: 250 VACrms (50 or 60 Hz) or less
FX1xxx-x-x-L: 60 VACrms (50 or 60 Hz) or less
Mounting position Can be inclined up to 30 degrees backward. Left and right horizontal.
Warm-up time At least 30 minutes after power on
Installation location Indoors
Operating altitude 2000 m or less
Power Supply
Item Specifications
Rated supply voltage 100 to 240 VAC
Allowable power supply voltage range
90 to 132, 180 to 264 VAC
Rated power supply frequency
50 Hz, 60 Hz
Power consumption Supply voltage LCD backlight off Normal Maximum
100 VAC 10 VA 15 VA 35 VA
240 VAC 15 VA 20 VA 45 VA
Allowable interruption time Less than 1 cycle of the power supply frequency