Manual
10-16
IM 04L21B01-01EN
Warning Messages
Code Message Ref. Section
600 Measured data and Settings have been initialized. –
601 Measured data have been initialized. –
610 This username is already registered. See section 8.2.
614 Calibration settings are reset because of range setting change. See section 3.9.
System Errors
Servicing is required when a system error occurs. If this happens, contact your nearest
YOKOGAWA dealer.
Code Message Description
901 ROM failure. −
902 RAM failure. −
910 A/D memory failure for all input channels. −
915 Failed to measure power. The FX temporarily failed to measure one of the elements
used to calculate the power.
916 The power measurement function failed. This is displayed when a malfunction is detected in the
power measurement section.
The computation icon is displayed in red.
921 Channel 1 A/D calibration value error. −
925 A/D calibration error. −
930 Memory acquisition failure. −
940 The Ethernet module is down. −
950 Incorrect number for the A/D calibration. −
951 Failed to write A/D calibration value. −