Manual
6-3
IM 04L21B01-01EN
Saving and Loading Data
3
2
1
4
5
6
7
8
9
10
11
12
13
14
App
Index
• Event data (when recording event data)
• Sample rate
Select the data recording interval. Use the table under “Data length” for reference.
• Mode
Settings Description
Free Records data continuously.
Single Records data when the trigger condition is met.
Repeat Records data each time the trigger condition is met.
• Data length
Select the size of a record data le. The recorded data is divided by the le size
specied here. The available data lengths vary depending on the number of
memory sampling channels and the Sample rate setting.
Sample rate
1
125 ms 250 ms 500 ms 1 s 2 s
Selectable range
of data length
10 min to
1 day
10 min to
2 days
10 min to
3 days
10 min to
7 days
10 min to
14 days
Sample rate
1
5 s 10 s 30 s 1 min 2 min
Selectable range
of data length
10 min to
31 days
10 min to
31 days
1 hour to
31 days
1 hour to
31 days
1 hour to
31 days
Sample rate
1
5 min 10 min
Selectable range
of data length
1 hour to
31 days
1 hour to
31 days
1 You cannot choose a sample rate that is higher than the scan interval.
• Pre-trigger
Specify the range when recording data before the trigger condition is met. Select
the range as a percentage of the data length from 0, 5, 25, 50, 75, 95, and 100%.
If you do not want to record the data existing before the trigger condition is met,
select 0%.
• Trigger signal > Key
Select On if you want to activate the trigger using key operation.
Note
• Triggers can be applied using event action (see section 7.1).
• If the trigger condition is already met when you press START, recording starts.
6.1 Setting the Recording Conditions of the Measured Data