3.13 Using the Log Scale to Perform Measurements (/LG1 Option)
Note
Set the mantissa to 1.00.
The output voltage for pressure reading whose mantissa is 1.00 has a certain deviation.
For example, if the pressure reading is 1.00E-5, the output voltage is 2.00 to 2.10 V. In this
situation, set the range to the minimum value (2.00 V).
Setting the Nonlinear Log Input (LogType1)
To support the nonlinear output of vacuum gauges, use the calibration correction
function (/CC1) to correct the recorder input values with linear approximation.
Conceptual Diagram
The pressure scale is mapped to an input scale that has been multiplied by a constant
number to derive the output scale.
10
9
8
7
6
5
4
3
2
1
0 1 2 3 4 5 6 7 8 9 10
10
-1
10
0
10
1
0
10
2
10
3
Input value [V]
Pressure scale
Output scale
Output value [V]
1
Press MENU (to switch to setting mode), and select the Menu tab > Meas channel
> Range, Alarm.
Set the range as shown below. To use the calibration correction function (/CC1
option), be sure to set the mode to LogType1.
Mode: LogType1
Range: 20V
Span: Lower 0.000, Upper 10.000
Scale: Lower 1.00E-1, Upper 4.00E+3
For the span and scale, assign the boundary values of the output scale and pressure
scale in the conceptual diagram.