Operation Manual

7-10
IM 1F5B2-01E
7.5 Self-checking
7.5.1 Checking for Failures
(1) Checking for Failures Using the BT200
Verify that:
• BT200 is connected correctly;
• BT200 is operating correctly; and
• Parameters are configured correctly;
while referring to the following examples of failures.
Connect the BT200 to
ULTRA YEWFLO and
press the key.
After the screen on the left
appears, press the
key.
ENTER
Example 1: Failure in Connection
FEED
Check cable connection.
Press the ENTER key.
BRAIN terminal
ID :
UTEL
ON/OFF
Communication failure.
Since communication is
not possible if the BT200 is
improperly connected to
ULTRA YEWFLO, the
message on the left
appears. Recheck the
connection. After
rechecking, press the
key (ESC).
F4
Example 2: Failure in Configuration of Parameters
OK
The Default Data screen
shows the current results
of a self-check on ULTRA
YEWFLO.
Pressing the key
(DIAG) on the Parameters
screen enters the self-
check screen “x60: SELF
CHECK.”
ESC
F2
Parameters
01 : MODEL
ULTRA YF
02 : TAG NO.
FIC–01
03 : SELF CHECK
ERROR
DIAG
DATA
PRNT
Parameters
B10 : TAG NO.
FIC–01
B20 : SIZE
25mm
B21 : PRESS RATING
L PRESS
ESC
FEED
PRNT
Self-check
B60 : SELF CHECK
ERROR
< ERROR >
< SPAN SET ERROR >
The Self-check screen
shows an error message if
there is an error.
ESC
(2) Checking for Failures Using the Indicator/Totalizer Component
NOTE
If any failure is found as a result of self-checking, the display of the indicator/totalizer
component shows an error number. If there is more than one failure, the display
changes from one error number to another at two-second intervals. For details on the
meanings of error numbers, see Table 7.5.1.
Figure 7.5-1 Checking for Failures Using the Indicator/Totalizer Component