Operation Manual

4-9
IM 1F5B2-01E
4. MOUNTING
Direction of flow
Nut
Collar
Gasket
Flange
Stud bolt
Figure 4.2-3 Horizontal Mounting
Vertical Mounting
<1> Thread one collars each through each of the lower two bolts as shown in fig 4.2-
4.
<2> Make sure, at that point, that both edges of ULTRA YEWFLO’s brim are in
proper contact with each pair of collars.
<3> Evenly fasten the bolts and nuts, tightening the edges onto the collars.
<4> Make sure the process fluid is not leaking.
Direction of flow
Figure 4.2-4 Vertical Mounting