Figure 4.2-1 Gasket with Pre-cut Bolt Holes (for Four-hole Flanges)
Flange of joint pipe
Joint pipe
Figure 4.2-2 Installation of Gaskets
IMPORTANT
• Use gaskets whose inner diameter is greater than those of ULTRA YEWFLO and the
joint pipe. Exercise extreme care to prevent the gaskets from protruding into the flow
path.
■ The procedure for mounting a wafer type of ULTRA YEWFLO is the same no matter which type of
flange you use.
Horizontal Mounting
<1> Thread two collars each through each of the lower two bolts as shown in fig 4.2-
3.
<2> Evenly fasten the nuts and bolts, tightening both edges of ULTRA YEWFLO’s
brim onto each pair of collars.
<3> Make sure the process fluid is not leaking.