User guide
2-9
IM 04L51B01-17EN
Commands and Responses
1
2
App
2.4 Setting Commands
SScan
Scan Interval
Sets the scan interval.
Syntax
SScan,p1,p2
p1
Scan group (1)
p2
Scan interval (100ms, 200ms, 500ms, 1s,
2s, 5s)
Query
SScan[,p1]?
Example Set the scan interval to 1 second.
SScan,1,1s
Description
• You cannot use this command to configure settings
while recording is in progress.
• You cannot use this command to configure settings
while computation is in progress.
SScanGroup
Scan Group
Registers a measurement channel in scan group 1.
Syntax
SScanGroup,p1,p2,p3
p1
Unit number (0)
p2
Module number (0 to 9)
p3
Scan group (1)
1
Scan group 1
Query
SScanGroup[,p1[,p2]]?
Example Set the module whose module number is 2 in
scan group 1.
SScanGroup,0,2,1
Description
• You cannot use this command to configure settings
while recording is in progress.
• You cannot use this command to configure settings
while computation is in progress.
SModeAI
AI Module
Sets the mode and A/D integration time of an AI module.
Syntax
SModeAI,p1,p2,p3,p4
p1
Unit number (0)
p2
Module number (0 to 9)
p3
Mode
2CH
2 channel mode
10CH
10 channel mode
p4
AD integration time (Auto, 50Hz, 60Hz,
Common)
Query
SModeAI[,p1[,p2]]?
Example For the module whose module number is 2, set
the mode to 10CH and the AD integration time
to Auto.
SModeAI,0,2,10CH,Auto
Description
• You cannot use this command to configure settings
while recording is in progress.
• You cannot use this command to configure settings
while computation is in progress.
• Scan intervals shorter than 1 s cannot be specified
if an electromagnetic relay scanner type (Type suffix
code: -T1) analog input module is in use (set up).
• Fixed to 10ch mode if an electromagnetic relay
scanner type analog input module is in use.
• You can set the parameters in the following
combinations.
Scan
Interval
Mode
(p3)
Integration time (p4)
Auto 50Hz 60Hz Common
100ms 2CH Yes Yes Yes No
10CH Yes No No No
200ms 2CH Yes Yes Yes No
10CH Yes No No No
500ms — Yes Yes Yes No
1s — Yes Yes Yes Yes
2s — Yes Yes Yes Yes
5s — Yes Yes Yes Yes
SModeDI
DI Module
Sets the mode of a DI module.
Syntax
SModeDI,p1,p2,p3
p1
Unit number (0)
p2
Module number (0 to 9)
p3
Mode (Normal, Remote)
Normal
DI input
Remote
Remote control input
Query
SModeDI[,p1[,p2]]?
Example Set the module whose module number is 2 as
a remote control input module.
SModeDI,0,2,Remote
Description
• You cannot use this command to configure settings
while recording is in progress.
• You cannot use this command to configure settings
while computation is in progress.
• Only one module can be set to remote. If different
modules are set to remote numerous times, the last
module will be the remote module.
SScaleOver
Detection of Values That Exceed the
Scale
Sets how to detect measurement over-range.
Syntax
SSclOver,p1
/P1
How to detect values that exceed the
scale
FREE
Assume scale over-range when the
measurement range is exceeded.
OVER
Assume scale over-range when
±105% of the scale is exceeded.
Query
SSclOver?
Example Assume scale over-range when the
measurement range is exceeded.
SSclOver,FREE