Owner's manual

2-43
IM MW100-17E
Response
1
2
3
AppApp
Index
Data Block Member Values
Member Name Value
Year 00 to 99 (70 to 99: 1970 to 1999, 00 to 69: 2000 to 2069)
Month 1 to 12
Day 1 to 31
Hours 0 to 23
Minute 0 to 59
Second 0 to 59
Milliseconds 0 to 990
Computation start 0, 1
Message information Bits 0:0, 1
Bits 1:0, 1
Setting key 0x0000 to 0xffff
Channel number 1 to 60: Measurement Channel
101 to 400: MATH channels
Alarm Status 0: No specification
1: Upper limit alarm setting
2: Lower limit alarm setting
3: Differential upper limit alarm setting
4: Differential lower limit alarm setting
5: Upper limit on rate-of-change alarm setting
6: Lower limit on rate-of-change alarm setting
7: Delay upper limit alarm setting
8: Delay lower limit alarm setting
Data value –32767 to 65535
–9999999 to 99999999
Unused Undefined
Special Data Values
During special statuses, measured and computed values are as shown in the table
below.
Special Data Value Type Value
+OVER 0x7fff 0x7fff
–OVER 0x8001 0x8001
Error 0x8004 0x8004
Note
Added information and alarm statuses are output without regard to the byte order specified
in the BO command.
Messages inside added information is uncertain.
2.3 Binary Output