Owner's manual

1-31
IM MW100-17E
Commands
1
2
3
AppApp
Index
SX Display Group Settings
Setting
SXp1,p2,p3<terminator>
p1 Display group number (01 to 18)
p2 Display group name (up to 16
characters)
p3 Channel structure (up to 100
characters, up to 20 channels)
Query
SX[p1]?
Example
Channel 1 and channels 3 to 6 are given a group name
of GROUP1 and set to group 1.
SX1,GROUP1,001.003-006
Set the channel structure by delimiting individual
channels with a period ( . ), and specifying ranges of
channels with a hyphen ( - ).
Explanation For the characters that can be used for group
names, see user specified characters in section 1.3.
See section 1.3 for the channel structure.
TL Trip Line Settings
Setting When p3=OFF
TLp1,p2,p3<terminator>
When p3=ON
TLp1,p2,p3,p4,p5<terminator>
p1
Display group number (01 to 18)
p2
Trip line number (1 to 4)
p3 Trip line display ON/OFF
p4 Display color (rrr.ggg.bbb)
rrr RGB R value (0 to 255)
ggg RGB G value (0 to 255)
bbb RGB B value (0 to 255)
p5 Display position (0 to 100)[%]
Query
TL[p1[,p2]]?
Example Display trip line 1 in group one in red at 10% position.
Set the top of the graph to 100% and the bottom to 0%.
TL1,1,ON,255.0.0,10
XR Channel and Tag Display
Switching
Setting
XRp1<terminator>
p1 Display type
CHANNEL Channel number display
TAG Tag display
Query
XR?
Example Set tag display.
XRTAG
1.5 Setting Commands (Display Related)
SG Message Settings
Setting
SGp1,p2<terminator>
p1 Message number (0 to 5)
p2 Message string (up to 15 characters)
Query
SG[p1]?
Example Set the MESSAGE1 string to message number 1.
SG1,MESSAGE1
Explanation For the characters that can be used for messages,
see user specified characters in section 1.3.
The message number 0 setting is valid in Setting
mode and Measurement mode.