Owner's manual

1-29
IM MW100-17E
Commands
1
2
3
AppApp
Index
XT Temperature Unit Settings
Setting
XTp1<terminator>
p1 Units of temperature
DEG_C Temperature calculated in °C
DEG_F Temperature calculated in °F
Query
XT?
Example Temperature units set in °F
XTDEG_F
Explanation When the temperature units are changed, the settings
of the SR, SF
, XB, XJ, XD, AO, AP, and ST commands
are initialized.
SW Daylight Saving Time Settings
Setting
SWp1,p2,p3,p4,p5<terminator>
p1 Starts/stops date/time type
START Start date/time
END Exit date/time
p2 Date/time-month
JAN January
FEB February
MAR March
APR April
MAY May
JUN June
JUL July
AUG August
SEP September
OCT October
NOV November
DEC December
p3 Date/time-week
1st First week
2nd Second week
3rd Third week
4th Fourth week
LAST Final week
p4 Date/time-day of week
SUN Sunday
MON Monday
TUE Tuesday
WED Wednesday
THU Thursday
FRI Friday
SAT Saturday
p5 Date/time-hour (0 to 23)
Query
SW[p1]?
Example Star Daylight Saving time on April, first week, Sunday, 2
hours.
SWSTART,APR,1st,SUN,2
SS Daylight Saving Time Operation
Settings
Setting
SSp1<terminator>
p1 Summer time operation valid/invalid
ON/OFF
Query
SS?
Example Enables Daylight Saving time.
SSON
BA Settings of Strain Input Initial
Balancing
Setting
BAp1,p2<terminator>
p1 Channel range (001 to 060)
p2 Initial balancing valid/invalid (ON/
OFF)
Query
BA[p1]?
Example Enable strain input initial balance operation on channel
1.
BA001,ON
Explanation
Specify measurement channels on p1.
UN Unit Number Settings
Setting
UNp1<terminator>
p1 Unit number (00 to 89)
Query
UN?
Example Set unit number to 5.
UN05
SV Rolling Average Settings
Setting When p2=OFF
SVp1,p2<terminator>
p1 MATH channel number (A001 to A060)
p2
Rolling average ON/OFF (OFF)
When p2=ON
SVp1,p2,p3,p4<terminator>
p1 MATH channel number (A001 to A060)
p2
Rolling average ON/OFF (ON)
p3 Sampling interval (1 to 6S, 10S, 12S,
15S, 20S, 30S, 1M to 6M, 10M, 12M,
15M, 20M, 30M, or 1H)
_S: Interval setting in unit of
seconds
_M: Interval setting in unit of
minutes
_H: Interval setting in unit of hours
p4 Number of samples (1 to 1500)
Query
SV[p1]?
Example
Set the sampling interval to 1 minute and the number of
samples to 360 on MATH channel A001.
SVA001,ON,1M,360
Explanation
This command can be used on models with the /M1
math option.
Set the sampling interval to an integer multiple of the
MATH interval.
1.4 Setting Commands (System Related)