Instruction Manual

1-42
IM MW100-01E
Thermocouple (cont.)
Measurement Range Notation Rated Measurement Range
Type-N N 0.0 to 1300.0°C
Type-W W 0.0 to 2315.0°C
Type-L L –200.0 to 900.0°C
Type-U U –200.0 to 400.0°C
KPvsAu7Fe KPvsAu7Fe 0.0 to 300.0 K
PLATINEL PLATINEL 0.0 to 1400.0°C
PR40-20 PR40-20 0.0 to 1900.0°C
NiNiMo NiNiMo 0.0 to 1310.0°C
WRe3-25 WRe3-25 0.0 to 2400.0°C
W/WRe26 WWRe26 0.0 to 2400.0°C
Type-N (AWG14) N14 0.0 to 1300.0°C
Type-XK GOST XK –200.0 to 600.0°C
DI
Measurement Range Notation Rated Measurement Range
LEVEL LEVEL Vth=2.4 V
Contact input CONTACT 1 kΩ or less, ON, 100 kΩ or less, OFF
(shunt capacitance: 0.01 μF or less)
Measurement Interval, Integration Time, and Filter
You can select from the following measurement intervals for this module.
500 ms, 1 s, 2 s, 5 s, 10 s, 20 s, 30 s, or 60 s.
The integral time and types of filters applied vary depending on the measurement
interval.
For information on filters, see section 2.9, “Measures Against Noise on the MW100 Data
Acquisition Unit.”
Measurement Synchronization
Since measurement is sequential, measurement on each channel is not synchronized.
1.6 Functions of the 30-CH, Medium-Speed DCV/TC/DI Input Module