6. Components
6.2 ZR402G Converter
Enter
Commissioning
r mA-output setup
r Alarms setup
r Contact setup
r Others
Basic setup
F7.6E.EPS
Complete Operation Display
d Interactive operations along with operation
display
d A variety of display modes2enabling you to
select the operation mode freely
d Back-lit LCD display allows viewing even in areas
of low lighting
d Error codes and details of errors are displayed no
need to refer to instruction manual
d Password for security
This data display provides for interactive operation.
d Example of basic display
d Example of trend display2displays data
changes
Error code Reason for error
E--1 Cell failure
E--2 Abnormal heater temperature
E--3 Defective A/D converter
E--4 Faulty EEPROM
ALARM1 Abnormal oxygen concentration
ALARM2 Abnormal moisture
ALARM3 Abnormal mixing ratio
ALARM6 Abnormal zero calibration factor
ALARM7 Abnormal span calibration factor
ALARM8 Stabilization time over
Typical Converter Displays
d Example of setting data display2displays data
changes
During automatic calibration,you can check has stabilized,
display data while viewing oxygen trend data,thus providing
highly reliable calibration.
d One-touch interactive display operation
d User-friendly design providing easy operation
without need to use the instruction manual
Self-testing suggests countermeasures for
problems
If a problem occurs, the liquid-crystal display will
provides an error code and description of the problem.
This enables prompt and appropriate corrective action
to be taken.
F6.3E.EPS
Figure 6.3 Converter