3. Installation
3.1.5 IECEx Flameproof Type
ZR22S-D Detector for use in hazardous locations:
Note 1: IECEx KEM 06.0006
IECEx flameproof type
Ex d IIB+ H
2
T2
Enclosure: IP66
IECEx type of protection "Dust"
Ex tD A21 IP66 T3008C
Enclosure: IP66
Note 2: Wiring
All wiring shall comply with local installation requirement.
The cable entry devices shall be of a certified flameproof type suitable for
the condition of use.
Note 3: Operation
Keep the "WARNING" label to the Detector.
WARNING: DO NOT OPEN WHEN ENERGIZED. INSTALL IN
ACCORDANCE WITH THE INSTRUCTION MANUAL.
USE AT LEAST 1508C HEAT RESISTANT CABLES.
Take care not to generate mechanical sparking when accessing to the
detector and peripheral devices in hazardous location.
Note 4: Maintenance and Repair
The detector modification or parts replacement by other than authorized
representative of Yokogawa Electric Corporation is prohibited and will void
IECEx Flameproof Certification.
Note 5: Cable Entry
The type of cable entry is marked near the cable entry port according to
following codes.
Type of Cable entry : Code
M20 3 1.5 : M
1/2NPT : A
In case of ANSI 1/2NPT plug, ANSI hexagonal wrench should be applied
to screw in.
The blanking elements shall be of certified flameproof type, suitable for the
conditions of use and correctly installed.
Certified cable glands that meet or exceed the requirements for Ex d IIB +
H
2
T2, Ex tD A21 IP66 T3008C, provide at least 6 threads engaged when
installed, and resist heat so that they can be used in the operating
environment, should be used.