Owner's manual
IM 11M13A01-02E
11-11
11. Inspection and Maintenance
To replace the fuse, follow these steps:
(1) Turn off the power to the converter for safe replacement.
(2) Remove the fuse from its holder. With the appropriate flat-blade screwdriver that just
fits the holder cap slot (Figure 11.7), turn the fuse holder cap 908 counterclockwise.
By doing so, you can remove the fuse together with the cap.
Cap
Flat-blade
screwdriver
Fuse
Socket
F11.7E.EPS
Figure 11.7 Removing the Fuse
(3) Check the rating of the fuse and that it satisfies the following:
Maximum rated voltage: 250 V
Maximum rated current: 3.15 amperes
Type: Time-lag fuse
Standards: UL-, CSA- or VDE-approved
Part number: A1113EF
Place a new, properly rated fuse in the holder together with the cap, and push and turn
the cap clockwise 90+‹with the screwdriver to complete installation of the fuse.
11.2.2 Cleaning
Use a soft dry cloth to clean any part of the converter during inspection and mainte-
nance.