Owner's manual
IM 11M13A01-02E
10-13
10. Other Functions
T10.5E-2.EPS
Item Data to be initialized Default setting
Parameter Oxygen
Alarm setting concentration
Hysteresis 0.1%O
2
Delayed action of alarm contact 3 seconds
Alarm data High-high alarm None
Alarm value 100%O
2
Alarm set value High-limit alarm None
Alarm value 100%O
2
Low-limit alarm None
Alarm value 0%O
2
Low-low alarm None
Alarm value 0%O
2
Alarm None
Warm-up On
Output range now being switched None
Now calibrating None
Contacts Output contact 1 Other settings Now maintenance servicing On
Blowback None
High-limit temp. alarm None
Calibration gas press. drop None
Unburnt gas detection None
Contact output action Open
Alarm None
Warm-up None
Output range switching None
Now calibrating On
Now maintenance servicing None
Output contact 2 Blowback None
High-limit temp. alarm None
Other settings Calibration gas press. drop None
Unburnt gas detection None
Contact output action Closed
High-high alarm None
High-limit alarm On
Output contact 3 Alarm Low-limit alarm On
Low-low alarm None
Calibration coefficient larm None
Startup power stabilization timeout None
Error None
Other settings None
Contact output Closed
Input contact 1 Function None
Input contact 2 Action Closed
Average value/ Average-value calculation One hour
Maximum and Intervals over which max. and min. 24 hours
minimum values values are monitored
Other data Amount of water vapor in exhaust gas 0.00 m
3
/kg (m
3
)
Fuel set value Theoretical amount of air 1.00 m
3
/kg (m
3
)
X value 1.00
Absolute humidity of atmosphere 0.0 kg/kg
Password Deleted