Owner's manual

IM 11M13A01-02E
10-12
Table 10.5 Initialization Items and Default Values
T10.5E-1.EPS
Item Data to be initialized Default setting
Type of equipment Not initialized
Equipment Detector ZR22
selection Measurement gas Wet gas
Display item 1st display item Oxygen concentration
2nd display item Current output 1
3rd display item Current output 2
Tag name Deleted
Trend graph Parameter Oxygen concentration
Sampling interval 30 seconds
Displayed data Upper limit (graph) 25%O
2
Lower limit (graph) 0%O
2
Automatic return time 0 min.
Language Not initialized
Mode Manual
Calibration procedure Span - zero
Zero-gas concentration 1.00%O
2
Span-gas concentration 21.00%O
2
Calibration data Calibration setting Output hold time 10 min., 00 sec.
Calibration time 10 min., 00 sec.
Interval 30 days, 00 hr.
Start date 01/01/00
Start time 00:00
Mode No function (invalid)
Blowback Blowback setting (Output) hold time 10 min., 00 sec.
Blowback time 10 min., 00 sec.
Interval 30 days, 00 hr.
Start date 01/01/00
Start time 00:00
Parameter Oxygen concentration
mA-output 1 Min. oxygen concentration 0%O
2
mA-output 2 Max. oxygen concentration 25%O
2
@@
Output damping 0%
Output mode Linear
Warm-up 4 mA
Set value 4 mA
Current output Maintenance Previous value held
data Output hold setting Set value 4 mA
Calibration, blowback Previous value held
Set value 4 mA
Error Preset value held
Set value 4 mA