Instruction Manual

IM 11M12A01-03E
10-15
10. Other Functions
Item Data to be initialized Default setting
Parameter Oxygen
Alarm setting concentration
Hysteresis 0.1% O
2
Delayed action of alarm contact 3 seconds
High-high alarm None
Alarm value 100% O
2
Alarm set value High-limit alarm None
Alarm value 100% O
2
Low-limit alarm None
Alarm value 0% O
2
Low-low alarm None
Alarm value 0% O
2
Humidity alarm Hysteresis 0.1% H
2
O
setting Contact delay 3 seconds
Alarm data High high alarm Off
Humidity alarm Alarm set value 100% H
2
O
set value High alarm Off
Alarm set value 100% H
2
O
Low alarm Off
Alarm set value 0% H
2
O
Low low alarm Off
Alarm set value 0% H
2
O
Mixing-ratio Hysteresis 0.01 kg/kg
alarm setting Contact delay 3 seconds
Mixing-ratio High high alarm Off
alarm set value Alarm set value 1 kg/kg
High alarm Off
Alarm set value 1 kg/kg
Low low alarm Off
Alarm set value 0 kg/kg
Low low alarm Off
Alarm set value 0 kg/kg
T10.5E-2.EPS
to be continued next page