Instruction Manual
IM 11M12A01-03E
10-15
10. Other Functions
 Item Data to be initialized Default setting
 Parameter    Oxygen
 Alarm setting    concentration
 Hysteresis   0.1% O
2
 Delayed action of alarm contact    3 seconds
     High-high alarm    None
   Alarm value    100% O
2
 Alarm set value High-limit alarm    None
 Alarm value    100% O
2
 Low-limit alarm    None
 Alarm value    0% O
2
 Low-low alarm    None
 Alarm value    0% O
2
 Humidity alarm Hysteresis  0.1% H
2
O
 setting  Contact delay 3  seconds
 Alarm data  High high alarm  Off
 Humidity alarm  Alarm set value  100% H
2
O
   set value  High alarm  Off
   Alarm set value  100% H
2
O
 Low alarm  Off
   Alarm set value  0% H
2
O
   Low low alarm  Off
   Alarm set value  0% H
2
O
 Mixing-ratio   Hysteresis  0.01 kg/kg
   alarm setting  Contact delay  3 seconds
 Mixing-ratio   High high alarm  Off
 alarm set value  Alarm set value  1 kg/kg
 High alarm  Off
   Alarm set value  1 kg/kg
   Low low alarm  Off
   Alarm set value  0 kg/kg
   Low low alarm  Off
   Alarm set value  0 kg/kg  
T10.5E-2.EPS
to be continued next page