Instruction Manual
IM 11M12A01-03E
10-14
Table 10.5 Initialization Items and Default Values
T10.5E-1.EPS
Item Data to be initialized Default setting
Type of equipment Not initialized
Detector ZR22
Measurement gas Wet gas
Display item 1st display item Humidity
2nd display item Current output 1
3rd display item Current output 2
Tag name Deleted
Displayed data Trend graph Parameter Humidity
Sampling interval 30 seconds
Upper limit (graph) 25% O
2
Lower limit (graph) 0% O
2
Automatic return time 0 min.
Language Not initialized
Mode Manual
Calibration procedure Span - zero
Zero-gas concentration 1.00% O
2
Span-gas concentration 21.00% O
2
Calibration data Calibration setting Output hold time 10 min., 00 sec.
Calibration time 10 min., 00 sec.
Interval 30 days, 00 hr.
Start date 10/01/00
Start time 00:00
Mode No function (invalid)
Blow back Blow back setting (Output) hold time 10 min., 00 sec.
Blow back time 10 min., 00 sec.
Interval 30 days, 00 hr.
Start date 01/01/00
Start time 00:00
Parameter Humidity
mA-output 1 Min. oxygen concentration 0% O
2
mA-output 2 Max. oxygen concentration 25% O
2
Output damping 0%
Output mode Linear
Warm-up 4 mA
Set value 4 mA
Maintenance Previous value held
Output hold setting Set value 4 mA
Calibration, blowback Previous value held
Set value 4 mA
Error Preset value held
Set value 3.4 mA
Equipment
selection
Current output
data
to be continued next page