Instruction Manual

IM 12D7J4-E/D/F-H
1
ENGLISH
1. GENERAL
1-1. Dimensions
Installation dimensions and basic construction details of the conductivity cells are shown in
fig.1.
1-2. Type number
The type number of the conductivity cell, shown on the cable, gives coded information on
the cell constant, the materials of construction etc.
1-3. Specifications
General specifications
Wetted parts materials
a. body & electrodes : Stainless steel AISI 316 or Titanium
b. insulations : PEEK
c. mounting adapter : PVDF or Stainless steel
Operating Specifications (Sensor)
Measuring system : 2-electrode
Maximum pressure : 10 Bar
Maximum temperature : 110° C
Temperature response : < 1 minute for 90% of a step change
Sterilize : At 135° C
Operating Specifications (Adapters)
See fig.1.0
0 10 20 30 40 50 60 70 80 100 120 140 160
BAR
ºC
2
4
6
8
10
12
SS
PVDF
Figure 1-0. Process/temperature relationship