Instruction Manual YOKOGAWA Model SC25V Combined 12mm sensor; pH, Ref, LE and Temperature IM12B6J1-40E-E 2nd edition
Table 1 Production year according to DIN. IEC 60062 Year 2010 2011 2012 2013 2014 2015 2016 2017 2018 2019 2020 2021 2022 IM12B6J1-40E-E Yearcode A B C D E F G H J K L M N Table 2 Production month according to DIN.
TABLE OF CONTENTS 1. Preface 1.1 Introduction 1.2 ATEX/IECEx compliance 1.3 Unpacking and checking 1.4 Warranty and service 4 4 4 4 4 2. General Specifications 2.1 Measuring elements 2.2 Construction materials 2.3 Functional specifications 2.4 Operating range 2.5 Dynamic specification 2.6 Sensor dimensions 2.7 Equipment and protective systems 5 5 5 5 5 5 6 6 3. Installation 3.1 Typical installation 3.2 Preparing the sensor for use 3.
1. PREFACE 1.1 Introduction This instruction manual provides information for the installation and use of the SC25V, combined 12mm glass sensor; pH, Ref, LE and Temperature. The SC25V is the best choice for the majority of typical processand waste water applications. The sensor is designed with titanium liquid earth, increased electrolyte reservoir and large surface area PTFE diaphragm combined with a long pathway inner junction resulting in a longtime stability and prolonged lifetime. 1.
2. GENERAL SPECIFICATIONS 2.1 Measuring elements pH glass electrode Silver/Silver Chloride reference Titanium liquid earth Pt1000 temperature sensor. 2.2 Construction materials Wetted materials: O-ring : Viton Earth Ring : Titanium Reference Junction : Porous PTFE Measuring Sensor : Glass Body Sensor : Glass, PEEK 2.
2.6 Sensor Dimensions Dimension according to DIN 19263:2007-05 Criteria : 120 mm version L (below plug head) : 120 ± 2 mm Ø shaft : 11.9 ± 0.1 mm Criteria : 225 mm version L (below plug head) : 225 ± 2 mm Ø shaft : 11.9 ± 0.1 mm Concentricity Perpendicularity Cilindricity 2.7 E quipment and protective systems intended for use in potentially explosive atmospheres ATEX II1G Ex ia IIC T3...T6 Ga.
3. INSTALLATION SC25V 3.1 Typical installation The SC25V sensor can be installed in-line, in a bypass loop or in an immersion assembly. For best results the SC25V should be mounted with the process fluid flowing towards the sensor and positioned at least 15° above the horizontal plane to eliminate air bubbles in the glass bulb. See fig. 6 section 3.3 3.2 Preparing the sensor for use Remove the sensor from its shipping box and slide off the “wet pocket” (the flask including the rubber ring).
Position 8 Position . 15° Minimum Figure 6 Figure 7 ø30 PG13.5 10 PG13,5 adapter K1523JA 20 5 ø30 ø36 Ø 54 PG 13,5 Conductivity Flow Fitting FF40 32 1/4” NPT Ø 36 Ø30 SC4A option /FF 100 (3.94”) 3/4” NPT 153 (6.0”) 231 60 PG13,5 adapter K1523JB 1/4” NPT 30 (1.18”) K1523JA Flow fitting FF40 with K1523JA Adapter to fit sensors with a PG13,5 process connection in FF40/FS40 and FD40 fittings. Material: Polypropylene. IM12B6J1-40E-E K1598AC Flow fitting option /FF K1598AC (incl. 3.
4. WIRING TO THE PH ANALYZER 4.1 Cable marking and pin allocation Signal description VP connector pH A pH-shield C Reference B Liquid earth D Temp -1 E Temp -2 F Cable Wire Colour (Using WU10-V-S) Transparent (core coax) Black Brown (shield coax) Yellow (cable shield) Red Blue 4.
4.5 Sensor VP connection VP-Connector (Temp 2) F (Temp 1) E (LE) D (pH) A (Ref.) B Figure 8 5. GENERAL CALIBRATION AND MAINTENANCE PROCEDURE 5.1 Calibration for pH measurement To calibrate a pH sensor, at least two buffer solutions with known pH values are required. It is recommended to use one buffer solution with a value of pH=7. The second buffer solution should be either acidic (pH<7) or alkaline (pH>7), depending on the process value to be measured.
5.3 Chemical cleaning HCN The chemical cleaning system is effective in processes where deposits can be removed by a suitable solvent. Good cleaning effects can be obtained from periodically spraying of an electrode. EXA PH450G has a built-in wash timer with programmable wash time, interval time and relaxation time for automatic cleaning of the electrodes, preventing pollution of the pH sensitive parts. After washing, it is possible to check the response time of the electrodes.
6. MODEL AND SUFFIX CODES 6.1 SC25V sensor Model Suffix Code Option code Description SC25V Combined 12mm sensor: pH, Ref, LE, Temp. Equiped with Variopin connector Sensor type -AGP25 General purpose -ALP25 High temp: Chemical resistant. Sensor length -120 120 mm -225 225 mm 6.
CHEMICAL COMPATIBILITY CHART 13 Material Viton EPDM Ti PTFE (teflon) PEEK Glass Temp. °C 20 60 100 20 60 100 20 60 100 20 60 100 20 60 100 20 60 100 8. APPENDIX Organic acid Oxidizing agent Neutral salt Basic salt Acid Alkali salt Organic acid Introganic acid Conc.
EC DECLARATION OF CONFORMITY IM12B6J1-40E-E
APPLICATION DATA SHEET SC25V COMBINED PH SENSOR Customer Address Contact Telephone Fax Please complete a separate form for each process stream to be analyzed and return, marked for the attention of: 1. Process data Serial no.
Yokogawa has an extensive sales and distribution network. Please refer to the European website YOKOGAWA ELECTRIC CORPORATION World Headquarters 9-32, Nakacho 2-chome, Musashino-shi Tokyo 180-8750 Japan www.yokogawa.com YOKOGAWA ELECTRIC ASIA Pte. LTD. 5 Bedok South Road Singapore 469270 Singapore www.yokogawa.com/sg YOKOGAWA CORPORATION OF AMERICA 2 Dart Road Newnan GA 30265 USA www.yokogawa.com/us YOKOGAWA CHINA CO. LTD. 3F Tower D Cartelo Crocodile Building No.