<4. GETTING STARTED>
4-2
4.2 Communication Setting
ToactivateMODBUS,thefollowingsettingsarerequired.TheyarelocatedintheMODBUSCommunication
menu in Detailed Setup.
Donotturnoffthepowersupplyimmediatelyaftersetting.WhentheparametersaresavedtotheEEPROM,
theredundantprocessingisexecutedfortheimprovementofreliability.Ifthepoweristurnedoffwithin60
secondsaftersettingismade,themodiedparametersarenotsavedandthesettingsmayreturntothe
previousvalues.
Table 4.1 Communication Parameters
Function Parameter Description and Setting
Deviceaddress Deviceaddress Select: 1 to 247
SetthedeviceaddressforROTAMASSwherethe
host expects to communicate. Make sure that no other
deviceonthebususesthesameaddress!
Baud rate Baud rate Select: 1200, 2400, 4800, 9600, 19200, 38400, 57600,
76800, 115200
MakesurethathostandROTAMASSusethesame
baud rate for communication
Communication
settings
Data format Select: Even 1 stop bit,Odd1stopbit,None1stop
bit,None2stopbits
MakesurethathostandROTAMASSusethesame
communication settings.
Bus mode Modbus mode Select: RTU,ASCII
MakesurethathostandROTAMASScommunicatein
the same Modbus mode.
Modbus init
IMPORTANT
The MODBUS communication module must be
restartedafterchanginganyoftheabovesettings!
This can be done by either powering off and on the
deviceorbyexecutingthefunction“Modbusinit”.
Please note that the settings in the menu will
beupdatedafterinitializationoftheModbus
communication module.
IMPORTANT
ConnectingtheROTAMASSMODBUScommunicationtooltoabuswithitsexistinghostmaycause
communication data scrambles resulting in a functional disorder or a system failure.
IMPORTANT