Instruction Manual
7-2
IM 01R04B04-00E-E 8th edition March 01, 2011 -00
All Rights Reserved. Copyright © 2003, Rota Yokogawa
The ROTAMASS software is continuously improved. This leads to functions, which are different in
HART (DD version 0401) and display setting (software version from 1.08.00)
Write protect Select No
- - R/W -
Yes
Enable WRT 10 min Acc. To HART - - R/W X Disable
New Password ASCII 8 characters - - R/W X Space
Software Seal Select Break - - R/W -
Keep
Master reset Select Not Execute - - R/W X Not Execute
Execute
Key Status Select Enable - - R/W Enable
Inhibit
Language Select English - - R/W X English
German
French
Russian
Mass Flow Decimal - 0 to 5 Mass ow unit R - -
Volume Flow Decimal - 0 to 5 Vol ow unit R - -
Density Decimal - 0 to 5 Density unit R - -
Temperature Decimal - 0 to 5 Temperature unit R - -
F-Total ex Decimal - Auto Total ex unit R X - -
R-Total ex Decimal - Auto Total ex unit R X - -
D-Total ex Decimal - Auto Total ex unit R X - -
F-Total mass Decimal - Auto Total mass unit R X - -
R-Total mass Decimal - Auto Total mass unit R X - -
D-Total mass Decimal - Auto Total mass unit R X - -
F-Total vol Decimal - Auto Total volume unit R X - -
R-Total vol Decimal - Auto Total volume unit R X - -
D-Total vol Decimal - Auto Total volume unit R X - -
F-Total net Decimal - Auto Total net unit R X - -
R-Total net Decimal - Auto Total net unit R X - -
D-Total net Decimal - Auto Total net unit R X - -
Concentration meas Decimal 0 to 110 0 to 5
Concentration
meas unit
R - -
Net ow Decimal - 0 to 5 Mass ow unit R - -
F-Total energy Decimal - Auto Total unit energy R X - -
Velocity Decimal - Auto Velocity unit R - -
Drive gain Decimal 0 to 11 (typ.) 3 V R - -
Analog output 1 Decimal -
As sel.
Var.
As sel. Var. R - -
Analog output 1% Decimal - 2 % R - -