If remote converter RCCF31 / RCCR31 was ordered with option /NC (no combination), the customer must
set the parameters of the connected detector by himself.
In parameter Detailed Setup / Sensor model the sensor model must be selected. Then, the sensor model
depending parameters are set automatically to default value :
Qnom, Mass ow unit, Mass ow LRV, Mass ow URV, Vol ow unit, Vol ow LRV, Vol ow URV, Autozero
range, Autozero uctuation range.
In parameter Detailed Setup / Sensor constants the sensor constants from calibration certicate must be set.
If material of wetted parts is HC see instruction TI 01R04B04-03E-E attached to the converter.
In Parameter Basic setup / Tag or Basic setup / Long tag the tag number must be set if necessary.