The terminals for the power supply and for the I/O ports are in a separate terminal box as shown in the
gures below.
Cable connection to rack-mount converter RCCR31:
1. Loose the cable clamp on terminal board
2. Remove the 25 mm section of PVC outer sheathing from the cable and x the outer r
ing of the cable
with the cable clamp on terminal board. (see left gure below).
3. Connect the numbered leads to the terminals as shown in the right gure below.
4. Connect inner shields to terminal COM.
Detector Cable Converter
RCCS3 RCCY03 RCCR31
D+ 1 1 D+
D- 2 2 D-
S1+ 3 3 S1+
S1- 4 4 S1-
S2+ 5 5 S2+
S2- 6 6 S2-
TP1 7 7 TP1
TP2 8 8 TP2
TP3 9 9 TP3
Shields COM