Instruction Manual
4-14
IM 01R04B04-00E-E 8th edition March 01, 2011 -00
All Rights Reserved. Copyright © 2003, Rota Yokogawa
Cable connection to detector RCCS3 and eld-mount converter RCCF31:
1. Remove connection box cover detector and converter (see gure below).
2. Loosen the thread of cable gland and insert the cable into the cap and the clamp part. (See gure below.)Loosen the thread of cable gland and insert the cable into the cap and the clamp part. (See gure below.)(See gure below.)
3. Remove the PVC outer sheathing 25 mm from the end and fold back outer shield over the clamp part.Remove the PVC outer sheathing 25 mm from the end and fold back outer shield over the clamp part.
(See gure below.)
4. Tighten the cable gland.
5. Connect the numbered leads to the terminals as shown in the gure below.
6. Connect inner shields to terminal COM.
7. Close the connection box covers.